บริการสอบเทียบของสถาบันมาตรวิทยาแห่งชาติ

Thermometry Metrology

* แหล่งข้อมูล : รายการบริการสอบเทียบสถาบันมาตรวิทยาแห่งชาติ

Item No.

Code

Description

Range

Accuracy
/Uncertainty

Remark

Service

1

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

- 38.8344 ?C

0.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

2

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

0.01 ?C

0.2 mK

Comparison against NIMT's Fixed Points

Calibration Fee

3

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

29.7646 ?C

0.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

4

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

231.928 ?C

1 mK

Comparison against NIMT's Fixed Points

Calibration Fee

5

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

419.527 ?C

1.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

6

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

660.323 ?C

3.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

7

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

961.78 ?C

5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

8

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

-189.3442 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

9

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

- 38.8344 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

10

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC

0.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

11

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

29.7646 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

12

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

231.928 oC

1.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

13

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

419.527 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

14

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

15

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

961.78 oC

10 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

16

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-189.3442 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

17

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

- 38.8344 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

18

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC

0.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

19

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

29.7646 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

20

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

231.928 oC

1.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

21

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

419.527 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

22

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

23

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

961.78 oC

10 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

24

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

-189.3442 oC to 0.01 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

25

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

-38.8344 oC to 0.01oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

26

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 29.7646 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

27

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 231.928 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

28

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 419.527 oC

3 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

29

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

30

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 961.78 oC

15 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

31

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-189.3442 oC to 0.01 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

32

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-38.8344 oC to 0.01oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

33

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 29.7646 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

34

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 419.527 oC

3 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

35

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

36

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 961.78 oC

15 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

37

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

-196oC

0.15 K

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

38

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

-80oC to -60oC

40 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

39

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

>-60oC to -40oC

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

40

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

- 40 ?C to 0 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

41

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 0 ?C to 50 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

42

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 50 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

43

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 250 ?C to 420 ?C

30 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

44

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 420 ?C to 550 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

45

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

>550oC to 650oC

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

46

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-196oC

0.15 K

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

47

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-100oC to -80oC

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

48

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-80oC to -60oC

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

49

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

>-60oC to -40oC

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

50

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

- 40 ?C to 0 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

51

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 0 ?C to 50 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

52

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 50 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

53

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 250 ?C to 420 ?C

30 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

54

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 420 ?C to 550 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

55

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

>550oC to 660oC

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

56

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

231.928 ?C

0.1 K

By ITS-90 Fixed Point Cells

Calibration Fee

57

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

419.527 ?C

0.1 K

Minimum sheath length 500 mm

Calibration Fee

58

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

660.323 ?C

0.2 K

Minimum sheath length 500 mm

Calibration Fee

59

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

961.78 ?C

0.2 K

Minimum sheath length 500 mm

Calibration Fee

60

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

1084.62 oC

0.3 K

Minimum sheath length 500 mm

Calibration Fee

61

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

1324.0 oC

0.6 K

By Co-C fixed point cell Minimum sheath length 700 mm

Calibration Fee

62

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

0 ?C to 660 ?C

0.2 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

63

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

> 660 ?C to 1100 ?C

0.3 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

64

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

> 1100 ?C to 1324 ?C

0.6 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

65

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

-40 ?C to 0 ?C

0.2 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

66

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

0 ?C to 250 ?C

0.1 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

67

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>250 ?C to 500 ?C

0.2 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

68

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>500 ?C to 700 ?C

0.5 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

69

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>700 ?C to 1100 ?C

1.0 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

70

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>1100 ?C to 1324 ?C

1.5 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

71

11020 - 12321

Base Metal Thermocouples

-40 ?C to 1100 ?C

0.1 - 1.5 K

By comparison with standard TC or SPRT in comparison furnace

Calibration Fee

72

11020 - 12411

Liquid-in-glass Thermometer

-40 ?C to 90 ?C

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

73

11020 - 12411

Liquid-in-glass Thermometer

> 90 ?C to 110 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

74

11020 - 12411

Liquid-in-glass Thermometer

> 110 ?C to 250 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

75

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

-80oC to -40oC (Resolution < 0.01oC)

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

76

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

- 40 ?C to 90 ?C

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

77

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

> 90 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

78

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

> 250 ?C to 450 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

79

11020 - 12712

Digital Thermometer (with probe) (resolution>0.2K)

-80oC to -40oC (Resolution > 0.1oC)

0.1 K

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

80

11020 - 12712

Digital Thermometer (with probe) (resolution>0.2K)

- 40 ?C to 450 ?C

0.1 K

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

81

11020 - 22311

Supply of Thermocouple Type S with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

82

11020 - 22312

Supply of Thermocouple Type R with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

83

11020 - 22313

Supply of Thermocouple Type B with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

84

11020 - 22314

Fabrication of Standard Thermocouples Type R, Type S, Type B with fixed point calibration (wires not included)

0 ?C to 1100 ?C

0.3 ?C

Electrical Annealing and re-fabrication thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

85

11020 - 12861

Temperature Bath

-80oC to 25oC (Alcohol)

20 mK

By Comparison with SPRTs

Calibration Fee

86

11020 - 12861

Temperature Bath

>5 ?C to 50 ?C (Water)

15 mK

By Comparison with SPRTs

Calibration Fee

87

11020 - 12861

Temperature Bath

>40 ?C to 250 ?C (Oil)

30 mK

By Comparison with SPRTs

Calibration Fee

88

11020 - 12861

Temperature Bath

>300oC to 500oC (Salt)

50 mK

By Comparison with SPRTs

Calibration Fee

89

11020 - 12862

Dry Block and Furnace

- 40 ?C to 420 ?C

0.1 K

By Comparison with SPRTs

Calibration Fee

90

11020 - 12862

Dry Block and Furnace

>420oC to 660oC

0.1 K

By Comparison with SPRTs

Calibration Fee

91

11030 - 11121

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

1084.62oC

0.20 K

Comparison to Standard Fixed-point Blackbody using 0.65 ?m standard radiation thermometer

Calibration Fee

92

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

1 084.62oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

93

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

961.78oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

94

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

660.323oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

95

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

419.527oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

96

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

961.78oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

97

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

660.323oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

98

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

419.527oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

99

11030 - 11411

Optical Pyrometer

156.5985 ?C to 1084.60 ?C

0.3 K to 0.5 K

By Comparison with Black Body Fixed Points

Calibration Fee

100

11030 - 11412

0.9 ?m radiation thermometer

960 ?C to 2000 ?C

0.39 K to 2.4 K

Direct realisation by measurement of spectral responsivity

Calibration Fee

101

11030 - 11413

0.65 ?m radiation thermometer

960 ?C to 2500 ?C

0.28 K to 1.6 K

Direct realisation by measurement of spectral responsivity

Calibration Fee

102

11030-12521

Variable Temperature Blackbody Radiation Sources

0 ?C to < 15 ?C

0.4 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 13 mm

Calibration Fee

103

11030-12525

Variable Temperature Blackbody Radiation Sources

15 ?C to 500 ?C

0.4 K to 2.1 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 13 mm

Calibration Fee

104

11030-12526

Variable Temperature Blackbody Radiation Sources

-20 ?C to < 15 ?C

0.4 K to 0.2 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 14 mm

Calibration Fee

105

11030-12527

Variable Temperature Blackbody Radiation Sources

15 ?C to 500 ?C

0.2 K to 2.1 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 14 mm

Calibration Fee

106

11030-12522

Variable Temperature Blackbody Radiation Sources

150 ?C to 1 100 ?C

0.7 K to 1.6 K

Comparison with 1.6 mm standard radiation thermometer

Calibration Fee

107

11030-12523

Variable Temperature Blackbody Radiation Sources

1 000 ?C to 2 500 ?C

1.6 K to 2.8 K

Comparison with 650 nm standard radiation thermometer

Calibration Fee

108

11030 - 12524

Blackbody sources for infrared ear thermometers

35 ? C to 42 ?C

55 mK

Comparison with standard liquid bath blackbodies through infrared ear thermometers

Calibration Fee

109

11030 - 12541

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

-5 ?C to 15 ?C

0.5 K to 1.5 K

By Comparison with PRT or Transfer Standard Radiation Thermometer

Calibration Fee

110

11030 - 12542

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

>15 ?C to < 1100 ?C

0.5 K to 5.0 K

By Comparison with Transfer Standard Radiation Thermometer or PRT

Calibration Fee

111

11030 - 12543

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

1100 ?C to 2500 ?C

2 K to 5 K

By Comparison with Transfer Standard Radiation Thermometer

Calibration Fee

112

11030-12544

Infrared ear thermometer (Resolution ? 0.1 K)

35 ?C to 42 ?C

0.2 K

Comparison with PRT in liquid bath blackbody

Calibration Fee

113

11030-12545

Infrared ear thermometer (Resolution < 0.1 K)

35 ?C to 42 ?C

0.10 K

Comparison with PRT in liquid bath blackbody

Calibration Fee

114

11040 - 13111

Dew Point Hygrometer

- 75 ?C to -25 ?C

0.3 K

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

115

11040 - 13112

Dew Point Hygrometer

>- 25 ?C to +1 ?C

0.2 K

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

116

11040 - 13113

Dew Point Hygrometer

>1 ?C to 25 ?C

0.1 K

Comparison with two pressure Humidity Generator

Calibration Fee

117

11040 - 13114

Dew Point Hygrometer

>25 ?C to 30 ?C

0.2 K

Comparison with two pressure Humidity Generator

Calibration Fee

118

11040-13115

Dew-Point Hygrometer

>30 oC to 60 oC

0.1 oC

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

119

11040 - 13311

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

15 % to 50 %

0.5 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

120

11040 - 13312

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

> 50 % to 80 %

0.7 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

121

11040 - 13313

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

> 80 % to 98 %

1.0 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

122

11040 - 13411

Thermo-Hygrograph

Temperature 15 ?C to 50 ?C

(0.5 to 1) ?C

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

123

11040 - 13411

Thermo-Hygrometer

Humidity (35 to 95)%RH

(1.5 to 3) %RH

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

124

11040 - 13412

Data Logger (Air Type)

Temperature 15 ?C to 50 ?C

(0.5 to 1) ?C

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

125

11040 - 13412

Data Logger (Air Type)

Humidity (35 to 95)%RH

(1.5 to 3) %RH

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

126

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

- 38.8344 ?C

0.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

127

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

0.01 ?C

0.2 mK

Comparison against NIMT's Fixed Points

Calibration Fee

128

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

29.7646 ?C

0.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

129

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

231.928 ?C

1 mK

Comparison against NIMT's Fixed Points

Calibration Fee

130

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

419.527 ?C

1.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

131

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

660.323 ?C

3.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

132

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

961.78 ?C

5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

133

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

-189.3442 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

134

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

- 38.8344 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

135

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC

0.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

136

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

29.7646 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

137

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

231.928 oC

1.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

138

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

419.527 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

139

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

140

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

961.78 oC

10 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

141

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-189.3442 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

142

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

- 38.8344 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

143

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC

0.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

144

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

29.7646 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

145

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

231.928 oC

1.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

146

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

419.527 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

147

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

148

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

961.78 oC

10 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

149

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

-189.3442 oC to 0.01 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

150

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

-38.8344 oC to 0.01oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

151

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 29.7646 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

152

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 231.928 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

153

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 419.527 oC

3 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

154

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

155

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 961.78 oC

15 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

156

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-189.3442 oC to 0.01 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

157

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-38.8344 oC to 0.01oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

158

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 29.7646 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

159

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 419.527 oC

3 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

160

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

161

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 961.78 oC

15 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

162

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

-196oC

0.15 K

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

163

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

-80oC to -60oC

40 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

164

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

>-60oC to -40oC

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

165

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

- 40 ?C to 0 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

166

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 0 ?C to 50 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

167

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 50 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

168

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 250 ?C to 420 ?C

30 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

169

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 420 ?C to 550 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

170

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

>550oC to 650oC

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

171

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-196oC

0.15 K

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

172

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-100oC to -80oC

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

173

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-80oC to -60oC

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

174

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

>-60oC to -40oC

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

175

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

- 40 ?C to 0 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

176

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 0 ?C to 50 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

177

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 50 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

178

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 250 ?C to 420 ?C

30 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

179

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 420 ?C to 550 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

180

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

>550oC to 660oC

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

181

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

231.928 ?C

0.1 K

By ITS-90 Fixed Point Cells

Calibration Fee

182

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

419.527 ?C

0.1 K

Minimum sheath length 500 mm

Calibration Fee

183

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

660.323 ?C

0.2 K

Minimum sheath length 500 mm

Calibration Fee

184

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

961.78 ?C

0.2 K

Minimum sheath length 500 mm

Calibration Fee

185

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

1084.62 oC

0.3 K

Minimum sheath length 500 mm

Calibration Fee

186

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

1324.0 oC

0.6 K

By Co-C fixed point cell Minimum sheath length 700 mm

Calibration Fee

187

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

0 ?C to 660 ?C

0.2 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

188

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

> 660 ?C to 1100 ?C

0.3 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

189

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

> 1100 ?C to 1324 ?C

0.6 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

190

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

-40 ?C to 0 ?C

0.2 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

191

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

0 ?C to 250 ?C

0.1 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

192

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>250 ?C to 500 ?C

0.2 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

193

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>500 ?C to 700 ?C

0.5 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

194

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>700 ?C to 1100 ?C

1.0 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

195

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>1100 ?C to 1324 ?C

1.5 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

196

11020 - 12321

Base Metal Thermocouples

-40 ?C to 1100 ?C

0.1 - 1.5 K

By comparison with standard TC or SPRT in comparison furnace

Calibration Fee

197

11020 - 12411

Liquid-in-glass Thermometer

-40 ?C to 90 ?C

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

198

11020 - 12411

Liquid-in-glass Thermometer

> 90 ?C to 110 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

199

11020 - 12411

Liquid-in-glass Thermometer

> 110 ?C to 250 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

200

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

-80oC to -40oC (Resolution < 0.01oC)

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

201

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

- 40 ?C to 90 ?C

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

202

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

> 90 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

203

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

> 250 ?C to 450 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

204

11020 - 12712

Digital Thermometer (with probe) (resolution>0.2K)

-80oC to -40oC (Resolution > 0.1oC)

0.1 K

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

205

11020 - 12712

Digital Thermometer (with probe) (resolution>0.2K)

- 40 ?C to 450 ?C

0.1 K

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

206

11020 - 22311

Supply of Thermocouple Type S with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

207

11020 - 22312

Supply of Thermocouple Type R with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

208

11020 - 22313

Supply of Thermocouple Type B with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

209

11020 - 22314

Fabrication of Standard Thermocouples Type R, Type S, Type B with fixed point calibration (wires not included)

0 ?C to 1100 ?C

0.3 ?C

Electrical Annealing and re-fabrication thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

210

11020 - 12861

Temperature Bath

-80oC to 25oC (Alcohol)

20 mK

By Comparison with SPRTs

Calibration Fee

211

11020 - 12861

Temperature Bath

>5 ?C to 50 ?C (Water)

15 mK

By Comparison with SPRTs

Calibration Fee

212

11020 - 12861

Temperature Bath

>40 ?C to 250 ?C (Oil)

30 mK

By Comparison with SPRTs

Calibration Fee

213

11020 - 12861

Temperature Bath

>300oC to 500oC (Salt)

50 mK

By Comparison with SPRTs

Calibration Fee

214

11020 - 12862

Dry Block and Furnace

- 40 ?C to 420 ?C

0.1 K

By Comparison with SPRTs

Calibration Fee

215

11020 - 12862

Dry Block and Furnace

>420oC to 660oC

0.1 K

By Comparison with SPRTs

Calibration Fee

216

11030 - 11121

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

1084.62oC

0.20 K

Comparison to Standard Fixed-point Blackbody using 0.65 ?m standard radiation thermometer

Calibration Fee

217

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

1 084.62oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

218

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

961.78oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

219

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

660.323oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

220

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

419.527oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

221

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

961.78oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

222

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

660.323oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

223

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

419.527oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

224

11030 - 11411

Optical Pyrometer

156.5985 ?C to 1084.60 ?C

0.3 K to 0.5 K

By Comparison with Black Body Fixed Points

Calibration Fee

225

11030 - 11412

0.9 ?m radiation thermometer

960 ?C to 2000 ?C

0.39 K to 2.4 K

Direct realisation by measurement of spectral responsivity

Calibration Fee

226

11030 - 11413

0.65 ?m radiation thermometer

960 ?C to 2500 ?C

0.28 K to 1.6 K

Direct realisation by measurement of spectral responsivity

Calibration Fee

227

11030-12521

Variable Temperature Blackbody Radiation Sources

0 ?C to < 15 ?C

0.4 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 13 mm

Calibration Fee

228

11030-12525

Variable Temperature Blackbody Radiation Sources

15 ?C to 500 ?C

0.4 K to 2.1 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 13 mm

Calibration Fee

229

11030-12526

Variable Temperature Blackbody Radiation Sources

-20 ?C to < 15 ?C

0.4 K to 0.2 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 14 mm

Calibration Fee

230

11030-12527

Variable Temperature Blackbody Radiation Sources

15 ?C to 500 ?C

0.2 K to 2.1 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 14 mm

Calibration Fee

231

11030-12522

Variable Temperature Blackbody Radiation Sources

150 ?C to 1 100 ?C

0.7 K to 1.6 K

Comparison with 1.6 mm standard radiation thermometer

Calibration Fee

232

11030-12523

Variable Temperature Blackbody Radiation Sources

1 000 ?C to 2 500 ?C

1.6 K to 2.8 K

Comparison with 650 nm standard radiation thermometer

Calibration Fee

233

11030 - 12524

Blackbody sources for infrared ear thermometers

35 ? C to 42 ?C

55 mK

Comparison with standard liquid bath blackbodies through infrared ear thermometers

Calibration Fee

234

11030 - 12541

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

-5 ?C to 15 ?C

0.5 K to 1.5 K

By Comparison with PRT or Transfer Standard Radiation Thermometer

Calibration Fee

235

11030 - 12542

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

>15 ?C to < 1100 ?C

0.5 K to 5.0 K

By Comparison with Transfer Standard Radiation Thermometer or PRT

Calibration Fee

236

11030 - 12543

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

1100 ?C to 2500 ?C

2 K to 5 K

By Comparison with Transfer Standard Radiation Thermometer

Calibration Fee

237

11030-12544

Infrared ear thermometer (Resolution ? 0.1 K)

35 ?C to 42 ?C

0.2 K

Comparison with PRT in liquid bath blackbody

Calibration Fee

238

11030-12545

Infrared ear thermometer (Resolution < 0.1 K)

35 ?C to 42 ?C

0.10 K

Comparison with PRT in liquid bath blackbody

Calibration Fee

239

11040 - 13111

Dew Point Hygrometer

- 75 ?C to -25 ?C

0.3 K

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

240

11040 - 13112

Dew Point Hygrometer

>- 25 ?C to +1 ?C

0.2 K

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

241

11040 - 13113

Dew Point Hygrometer

>1 ?C to 25 ?C

0.1 K

Comparison with two pressure Humidity Generator

Calibration Fee

242

11040 - 13114

Dew Point Hygrometer

>25 ?C to 30 ?C

0.2 K

Comparison with two pressure Humidity Generator

Calibration Fee

243

11040-13115

Dew-Point Hygrometer

>30 oC to 60 oC

0.1 oC

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

244

11040 - 13311

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

15 % to 50 %

0.5 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

245

11040 - 13312

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

> 50 % to 80 %

0.7 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

246

11040 - 13313

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

> 80 % to 98 %

1.0 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

247

11040 - 13411

Thermo-Hygrograph

Temperature 15 ?C to 50 ?C

(0.5 to 1) ?C

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

248

11040 - 13411

Thermo-Hygrometer

Humidity (35 to 95)%RH

(1.5 to 3) %RH

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

249

11040 - 13412

Data Logger (Air Type)

Temperature 15 ?C to 50 ?C

(0.5 to 1) ?C

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

250

11040 - 13412

Data Logger (Air Type)

Humidity (35 to 95)%RH

(1.5 to 3) %RH

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

251

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

- 38.8344 ?C

0.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

252

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

0.01 ?C

0.2 mK

Comparison against NIMT's Fixed Points

Calibration Fee

253

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

29.7646 ?C

0.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

254

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

231.928 ?C

1 mK

Comparison against NIMT's Fixed Points

Calibration Fee

255

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

419.527 ?C

1.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

256

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

660.323 ?C

3.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

257

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

961.78 ?C

5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

258

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

-189.3442 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

259

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

- 38.8344 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

260

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC

0.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

261

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

29.7646 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

262

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

231.928 oC

1.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

263

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

419.527 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

264

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

265

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

961.78 oC

10 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

266

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-189.3442 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

267

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

- 38.8344 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

268

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC

0.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

269

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

29.7646 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

270

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

231.928 oC

1.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

271

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

419.527 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

272

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

273

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

961.78 oC

10 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

274

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

-189.3442 oC to 0.01 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

275

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

-38.8344 oC to 0.01oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

276

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 29.7646 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

277

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 231.928 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

278

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 419.527 oC

3 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

279

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

280

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 961.78 oC

15 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

281

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-189.3442 oC to 0.01 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

282

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-38.8344 oC to 0.01oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

283

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 29.7646 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

284

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 419.527 oC

3 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

285

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

286

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 961.78 oC

15 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

287

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

-196oC

0.15 K

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

288

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

-80oC to -60oC

40 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

289

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

>-60oC to -40oC

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

290

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

- 40 ?C to 0 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

291

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 0 ?C to 50 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

292

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 50 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

293

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 250 ?C to 420 ?C

30 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

294

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 420 ?C to 550 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

295

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

>550oC to 650oC

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

296

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-196oC

0.15 K

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

297

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-100oC to -80oC

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

298

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-80oC to -60oC

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

299

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

>-60oC to -40oC

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

300

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

- 40 ?C to 0 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

301

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 0 ?C to 50 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

302

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 50 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

303

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 250 ?C to 420 ?C

30 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

304

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 420 ?C to 550 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

305

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

>550oC to 660oC

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

306

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

231.928 ?C

0.1 K

By ITS-90 Fixed Point Cells

Calibration Fee

307

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

419.527 ?C

0.1 K

Minimum sheath length 500 mm

Calibration Fee

308

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

660.323 ?C

0.2 K

Minimum sheath length 500 mm

Calibration Fee

309

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

961.78 ?C

0.2 K

Minimum sheath length 500 mm

Calibration Fee

310

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

1084.62 oC

0.3 K

Minimum sheath length 500 mm

Calibration Fee

311

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

1324.0 oC

0.6 K

By Co-C fixed point cell Minimum sheath length 700 mm

Calibration Fee

312

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

0 ?C to 660 ?C

0.2 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

313

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

> 660 ?C to 1100 ?C

0.3 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

314

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

> 1100 ?C to 1324 ?C

0.6 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

315

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

-40 ?C to 0 ?C

0.2 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

316

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

0 ?C to 250 ?C

0.1 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

317

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>250 ?C to 500 ?C

0.2 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

318

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>500 ?C to 700 ?C

0.5 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

319

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>700 ?C to 1100 ?C

1.0 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

320

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>1100 ?C to 1324 ?C

1.5 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

321

11020 - 12321

Base Metal Thermocouples

-40 ?C to 1100 ?C

0.1 - 1.5 K

By comparison with standard TC or SPRT in comparison furnace

Calibration Fee

322

11020 - 12411

Liquid-in-glass Thermometer

-40 ?C to 90 ?C

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

323

11020 - 12411

Liquid-in-glass Thermometer

> 90 ?C to 110 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

324

11020 - 12411

Liquid-in-glass Thermometer

> 110 ?C to 250 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

325

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

-80oC to -40oC (Resolution < 0.01oC)

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

326

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

- 40 ?C to 90 ?C

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

327

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

> 90 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

328

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

> 250 ?C to 450 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

329

11020 - 12712

Digital Thermometer (with probe) (resolution>0.2K)

-80oC to -40oC (Resolution > 0.1oC)

0.1 K

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

330

11020 - 12712

Digital Thermometer (with probe) (resolution>0.2K)

- 40 ?C to 450 ?C

0.1 K

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

331

11020 - 22311

Supply of Thermocouple Type S with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

332

11020 - 22312

Supply of Thermocouple Type R with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

333

11020 - 22313

Supply of Thermocouple Type B with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

334

11020 - 22314

Fabrication of Standard Thermocouples Type R, Type S, Type B with fixed point calibration (wires not included)

0 ?C to 1100 ?C

0.3 ?C

Electrical Annealing and re-fabrication thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

335

11020 - 12861

Temperature Bath

-80oC to 25oC (Alcohol)

20 mK

By Comparison with SPRTs

Calibration Fee

336

11020 - 12861

Temperature Bath

>5 ?C to 50 ?C (Water)

15 mK

By Comparison with SPRTs

Calibration Fee

337

11020 - 12861

Temperature Bath

>40 ?C to 250 ?C (Oil)

30 mK

By Comparison with SPRTs

Calibration Fee

338

11020 - 12861

Temperature Bath

>300oC to 500oC (Salt)

50 mK

By Comparison with SPRTs

Calibration Fee

339

11020 - 12862

Dry Block and Furnace

- 40 ?C to 420 ?C

0.1 K

By Comparison with SPRTs

Calibration Fee

340

11020 - 12862

Dry Block and Furnace

>420oC to 660oC

0.1 K

By Comparison with SPRTs

Calibration Fee

341

11030 - 11121

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

1084.62oC

0.20 K

Comparison to Standard Fixed-point Blackbody using 0.65 ?m standard radiation thermometer

Calibration Fee

342

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

1 084.62oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

343

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

961.78oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

344

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

660.323oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

345

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

419.527oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

346

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

961.78oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

347

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

660.323oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

348

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

419.527oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

349

11030 - 11411

Optical Pyrometer

156.5985 ?C to 1084.60 ?C

0.3 K to 0.5 K

By Comparison with Black Body Fixed Points

Calibration Fee

350

11030 - 11412

0.9 ?m radiation thermometer

960 ?C to 2000 ?C

0.39 K to 2.4 K

Direct realisation by measurement of spectral responsivity

Calibration Fee

351

11030 - 11413

0.65 ?m radiation thermometer

960 ?C to 2500 ?C

0.28 K to 1.6 K

Direct realisation by measurement of spectral responsivity

Calibration Fee

352

11030-12521

Variable Temperature Blackbody Radiation Sources

0 ?C to < 15 ?C

0.4 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 13 mm

Calibration Fee

353

11030-12525

Variable Temperature Blackbody Radiation Sources

15 ?C to 500 ?C

0.4 K to 2.1 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 13 mm

Calibration Fee

354

11030-12526

Variable Temperature Blackbody Radiation Sources

-20 ?C to < 15 ?C

0.4 K to 0.2 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 14 mm

Calibration Fee

355

11030-12527

Variable Temperature Blackbody Radiation Sources

15 ?C to 500 ?C

0.2 K to 2.1 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 14 mm

Calibration Fee

356

11030-12522

Variable Temperature Blackbody Radiation Sources

150 ?C to 1 100 ?C

0.7 K to 1.6 K

Comparison with 1.6 mm standard radiation thermometer

Calibration Fee

357

11030-12523

Variable Temperature Blackbody Radiation Sources

1 000 ?C to 2 500 ?C

1.6 K to 2.8 K

Comparison with 650 nm standard radiation thermometer

Calibration Fee

358

11030 - 12524

Blackbody sources for infrared ear thermometers

35 ? C to 42 ?C

55 mK

Comparison with standard liquid bath blackbodies through infrared ear thermometers

Calibration Fee

359

11030 - 12541

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

-5 ?C to 15 ?C

0.5 K to 1.5 K

By Comparison with PRT or Transfer Standard Radiation Thermometer

Calibration Fee

360

11030 - 12542

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

>15 ?C to < 1100 ?C

0.5 K to 5.0 K

By Comparison with Transfer Standard Radiation Thermometer or PRT

Calibration Fee

361

11030 - 12543

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

1100 ?C to 2500 ?C

2 K to 5 K

By Comparison with Transfer Standard Radiation Thermometer

Calibration Fee

362

11030-12544

Infrared ear thermometer (Resolution ? 0.1 K)

35 ?C to 42 ?C

0.2 K

Comparison with PRT in liquid bath blackbody

Calibration Fee

363

11030-12545

Infrared ear thermometer (Resolution < 0.1 K)

35 ?C to 42 ?C

0.10 K

Comparison with PRT in liquid bath blackbody

Calibration Fee

364

11040 - 13111

Dew Point Hygrometer

- 75 ?C to -25 ?C

0.3 K

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

365

11040 - 13112

Dew Point Hygrometer

>- 25 ?C to +1 ?C

0.2 K

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

366

11040 - 13113

Dew Point Hygrometer

>1 ?C to 25 ?C

0.1 K

Comparison with two pressure Humidity Generator

Calibration Fee

367

11040 - 13114

Dew Point Hygrometer

>25 ?C to 30 ?C

0.2 K

Comparison with two pressure Humidity Generator

Calibration Fee

368

11040-13115

Dew-Point Hygrometer

>30 oC to 60 oC

0.1 oC

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

369

11040 - 13311

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

15 % to 50 %

0.5 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

370

11040 - 13312

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

> 50 % to 80 %

0.7 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

371

11040 - 13313

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

> 80 % to 98 %

1.0 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

372

11040 - 13411

Thermo-Hygrograph

Temperature 15 ?C to 50 ?C

(0.5 to 1) ?C

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

373

11040 - 13411

Thermo-Hygrometer

Humidity (35 to 95)%RH

(1.5 to 3) %RH

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

374

11040 - 13412

Data Logger (Air Type)

Temperature 15 ?C to 50 ?C

(0.5 to 1) ?C

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

375

11040 - 13412

Data Logger (Air Type)

Humidity (35 to 95)%RH

(1.5 to 3) %RH

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

376

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

- 38.8344 ?C

0.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

377

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

0.01 ?C

0.2 mK

Comparison against NIMT's Fixed Points

Calibration Fee

378

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

29.7646 ?C

0.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

379

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

231.928 ?C

1 mK

Comparison against NIMT's Fixed Points

Calibration Fee

380

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

419.527 ?C

1.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

381

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

660.323 ?C

3.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

382

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

961.78 ?C

5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

383

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

-189.3442 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

384

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

- 38.8344 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

385

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC

0.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

386

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

29.7646 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

387

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

231.928 oC

1.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

388

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

419.527 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

389

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

390

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

961.78 oC

10 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

391

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-189.3442 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

392

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

- 38.8344 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

393

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC

0.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

394

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

29.7646 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

395

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

231.928 oC

1.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

396

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

419.527 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

397

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

398

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

961.78 oC

10 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

399

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

-189.3442 oC to 0.01 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

400

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

-38.8344 oC to 0.01oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

401

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 29.7646 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

402

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 231.928 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

403

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 419.527 oC

3 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

404

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

405

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 961.78 oC

15 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

406

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-189.3442 oC to 0.01 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

407

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-38.8344 oC to 0.01oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

408

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 29.7646 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

409

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 419.527 oC

3 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

410

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

411

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 961.78 oC

15 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

412

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

-196oC

0.15 K

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

413

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

-80oC to -60oC

40 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

414

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

>-60oC to -40oC

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

415

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

- 40 ?C to 0 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

416

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 0 ?C to 50 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

417

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 50 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

418

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 250 ?C to 420 ?C

30 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

419

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 420 ?C to 550 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

420

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

>550oC to 650oC

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

421

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-196oC

0.15 K

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

422

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-100oC to -80oC

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

423

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-80oC to -60oC

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

424

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

>-60oC to -40oC

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

425

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

- 40 ?C to 0 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

426

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 0 ?C to 50 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

427

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 50 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

428

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 250 ?C to 420 ?C

30 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

429

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 420 ?C to 550 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

430

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

>550oC to 660oC

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

431

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

231.928 ?C

0.1 K

By ITS-90 Fixed Point Cells

Calibration Fee

432

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

419.527 ?C

0.1 K

Minimum sheath length 500 mm

Calibration Fee

433

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

660.323 ?C

0.2 K

Minimum sheath length 500 mm

Calibration Fee

434

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

961.78 ?C

0.2 K

Minimum sheath length 500 mm

Calibration Fee

435

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

1084.62 oC

0.3 K

Minimum sheath length 500 mm

Calibration Fee

436

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

1324.0 oC

0.6 K

By Co-C fixed point cell Minimum sheath length 700 mm

Calibration Fee

437

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

0 ?C to 660 ?C

0.2 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

438

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

> 660 ?C to 1100 ?C

0.3 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

439

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

> 1100 ?C to 1324 ?C

0.6 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

440

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

-40 ?C to 0 ?C

0.2 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

441

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

0 ?C to 250 ?C

0.1 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

442

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>250 ?C to 500 ?C

0.2 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

443

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>500 ?C to 700 ?C

0.5 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

444

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>700 ?C to 1100 ?C

1.0 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

445

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>1100 ?C to 1324 ?C

1.5 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

446

11020 - 12321

Base Metal Thermocouples

-40 ?C to 1100 ?C

0.1 - 1.5 K

By comparison with standard TC or SPRT in comparison furnace

Calibration Fee

447

11020 - 12411

Liquid-in-glass Thermometer

-40 ?C to 90 ?C

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

448

11020 - 12411

Liquid-in-glass Thermometer

> 90 ?C to 110 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

449

11020 - 12411

Liquid-in-glass Thermometer

> 110 ?C to 250 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

450

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

-80oC to -40oC (Resolution < 0.01oC)

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

451

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

- 40 ?C to 90 ?C

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

452

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

> 90 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

453

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

> 250 ?C to 450 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

454

11020 - 12712

Digital Thermometer (with probe) (resolution>0.2K)

-80oC to -40oC (Resolution > 0.1oC)

0.1 K

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

455

11020 - 12712

Digital Thermometer (with probe) (resolution>0.2K)

- 40 ?C to 450 ?C

0.1 K

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

456

11020 - 22311

Supply of Thermocouple Type S with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

457

11020 - 22312

Supply of Thermocouple Type R with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

458

11020 - 22313

Supply of Thermocouple Type B with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

459

11020 - 22314

Fabrication of Standard Thermocouples Type R, Type S, Type B with fixed point calibration (wires not included)

0 ?C to 1100 ?C

0.3 ?C

Electrical Annealing and re-fabrication thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

460

11020 - 12861

Temperature Bath

-80oC to 25oC (Alcohol)

20 mK

By Comparison with SPRTs

Calibration Fee

461

11020 - 12861

Temperature Bath

>5 ?C to 50 ?C (Water)

15 mK

By Comparison with SPRTs

Calibration Fee

462

11020 - 12861

Temperature Bath

>40 ?C to 250 ?C (Oil)

30 mK

By Comparison with SPRTs

Calibration Fee

463

11020 - 12861

Temperature Bath

>300oC to 500oC (Salt)

50 mK

By Comparison with SPRTs

Calibration Fee

464

11020 - 12862

Dry Block and Furnace

- 40 ?C to 420 ?C

0.1 K

By Comparison with SPRTs

Calibration Fee

465

11020 - 12862

Dry Block and Furnace

>420oC to 660oC

0.1 K

By Comparison with SPRTs

Calibration Fee

466

11030 - 11121

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

1084.62oC

0.20 K

Comparison to Standard Fixed-point Blackbody using 0.65 ?m standard radiation thermometer

Calibration Fee

467

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

1 084.62oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

468

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

961.78oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

469

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

660.323oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

470

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

419.527oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

471

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

961.78oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

472

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

660.323oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

473

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

419.527oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

474

11030 - 11411

Optical Pyrometer

156.5985 ?C to 1084.60 ?C

0.3 K to 0.5 K

By Comparison with Black Body Fixed Points

Calibration Fee

475

11030 - 11412

0.9 ?m radiation thermometer

960 ?C to 2000 ?C

0.39 K to 2.4 K

Direct realisation by measurement of spectral responsivity

Calibration Fee

476

11030 - 11413

0.65 ?m radiation thermometer

960 ?C to 2500 ?C

0.28 K to 1.6 K

Direct realisation by measurement of spectral responsivity

Calibration Fee

477

11030-12521

Variable Temperature Blackbody Radiation Sources

0 ?C to < 15 ?C

0.4 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 13 mm

Calibration Fee

478

11030-12525

Variable Temperature Blackbody Radiation Sources

15 ?C to 500 ?C

0.4 K to 2.1 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 13 mm

Calibration Fee

479

11030-12526

Variable Temperature Blackbody Radiation Sources

-20 ?C to < 15 ?C

0.4 K to 0.2 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 14 mm

Calibration Fee

480

11030-12527

Variable Temperature Blackbody Radiation Sources

15 ?C to 500 ?C

0.2 K to 2.1 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 14 mm

Calibration Fee

481

11030-12522

Variable Temperature Blackbody Radiation Sources

150 ?C to 1 100 ?C

0.7 K to 1.6 K

Comparison with 1.6 mm standard radiation thermometer

Calibration Fee

482

11030-12523

Variable Temperature Blackbody Radiation Sources

1 000 ?C to 2 500 ?C

1.6 K to 2.8 K

Comparison with 650 nm standard radiation thermometer

Calibration Fee

483

11030 - 12524

Blackbody sources for infrared ear thermometers

35 ? C to 42 ?C

55 mK

Comparison with standard liquid bath blackbodies through infrared ear thermometers

Calibration Fee

484

11030 - 12541

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

-5 ?C to 15 ?C

0.5 K to 1.5 K

By Comparison with PRT or Transfer Standard Radiation Thermometer

Calibration Fee

485

11030 - 12542

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

>15 ?C to < 1100 ?C

0.5 K to 5.0 K

By Comparison with Transfer Standard Radiation Thermometer or PRT

Calibration Fee

486

11030 - 12543

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

1100 ?C to 2500 ?C

2 K to 5 K

By Comparison with Transfer Standard Radiation Thermometer

Calibration Fee

487

11030-12544

Infrared ear thermometer (Resolution ? 0.1 K)

35 ?C to 42 ?C

0.2 K

Comparison with PRT in liquid bath blackbody

Calibration Fee

488

11030-12545

Infrared ear thermometer (Resolution < 0.1 K)

35 ?C to 42 ?C

0.10 K

Comparison with PRT in liquid bath blackbody

Calibration Fee

489

11040 - 13111

Dew Point Hygrometer

- 75 ?C to -25 ?C

0.3 K

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

490

11040 - 13112

Dew Point Hygrometer

>- 25 ?C to +1 ?C

0.2 K

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

491

11040 - 13113

Dew Point Hygrometer

>1 ?C to 25 ?C

0.1 K

Comparison with two pressure Humidity Generator

Calibration Fee

492

11040 - 13114

Dew Point Hygrometer

>25 ?C to 30 ?C

0.2 K

Comparison with two pressure Humidity Generator

Calibration Fee

493

11040-13115

Dew-Point Hygrometer

>30 oC to 60 oC

0.1 oC

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

494

11040 - 13311

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

15 % to 50 %

0.5 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

495

11040 - 13312

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

> 50 % to 80 %

0.7 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

496

11040 - 13313

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

> 80 % to 98 %

1.0 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

497

11040 - 13411

Thermo-Hygrograph

Temperature 15 ?C to 50 ?C

(0.5 to 1) ?C

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

498

11040 - 13411

Thermo-Hygrometer

Humidity (35 to 95)%RH

(1.5 to 3) %RH

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

499

11040 - 13412

Data Logger (Air Type)

Temperature 15 ?C to 50 ?C

(0.5 to 1) ?C

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

500

11040 - 13412

Data Logger (Air Type)

Humidity (35 to 95)%RH

(1.5 to 3) %RH

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

501

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

- 38.8344 ?C

0.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

502

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

0.01 ?C

0.2 mK

Comparison against NIMT's Fixed Points

Calibration Fee

503

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

29.7646 ?C

0.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

504

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

231.928 ?C

1 mK

Comparison against NIMT's Fixed Points

Calibration Fee

505

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

419.527 ?C

1.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

506

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

660.323 ?C

3.5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

507

11010 - 11111

Fixed point cells(Primary and Secondary fixed points)

961.78 ?C

5 mK

Comparison against NIMT's Fixed Points

Calibration Fee

508

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

-189.3442 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

509

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

- 38.8344 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

510

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC

0.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

511

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

29.7646 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

512

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

231.928 oC

1.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

513

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

419.527 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

514

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

515

11010 - 11321

Standard Platinum Resistance Thermometer (SPRTs)

961.78 oC

10 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

516

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-189.3442 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

517

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

- 38.8344 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

518

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC

0.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

519

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

29.7646 oC

0.8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

520

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

231.928 oC

1.5 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

521

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

419.527 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

522

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

523

11010 - 11322

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

961.78 oC

10 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

524

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

-189.3442 oC to 0.01 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

525

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

-38.8344 oC to 0.01oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

526

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 29.7646 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

527

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 231.928 oC

2 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

528

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 419.527 oC

3 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

529

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

530

11010 - 11323

Standard Platinum Resistance Thermometer (SPRTs)

0.01 oC to 961.78 oC

15 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

531

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-189.3442 oC to 0.01 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

532

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

-38.8344 oC to 0.01oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

533

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 29.7646 oC

1 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

534

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 419.527 oC

3 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

535

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 660.323 oC

8 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

536

11010 - 11324

Semi-Standard Platinum Resistance Thermometer or Platinum Resistance Thermometer

0.01 oC to 961.78 oC

15 mK

By ITS-90 Fixed Point Cells, Quartz sheath for temperature above Zn. Minimum sheath length 450 mm

Calibration Fee

537

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

-196oC

0.15 K

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

538

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

-80oC to -60oC

40 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

539

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

>-60oC to -40oC

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

540

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

- 40 ?C to 0 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

541

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 0 ?C to 50 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

542

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 50 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

543

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 250 ?C to 420 ?C

30 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

544

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

> 420 ?C to 550 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

545

11020 - 12221

Standard Platinum Resistance Thermometer (SPRTs)

>550oC to 650oC

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

546

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-196oC

0.15 K

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

547

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-100oC to -80oC

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

548

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

-80oC to -60oC

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

549

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

>-60oC to -40oC

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

550

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

- 40 ?C to 0 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

551

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 0 ?C to 50 ?C

10 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

552

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 50 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

553

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 250 ?C to 420 ?C

30 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

554

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

> 420 ?C to 550 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

555

11020 - 12222

Industrial Platinum Resistance Thermometers (IPRTs)

>550oC to 660oC

50 mK

By Comparison with SPRTs in stirred liquid bath Minimum sheath length 300 mm

Calibration Fee

556

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

231.928 ?C

0.1 K

By ITS-90 Fixed Point Cells

Calibration Fee

557

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

419.527 ?C

0.1 K

Minimum sheath length 500 mm

Calibration Fee

558

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

660.323 ?C

0.2 K

Minimum sheath length 500 mm

Calibration Fee

559

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

961.78 ?C

0.2 K

Minimum sheath length 500 mm

Calibration Fee

560

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

1084.62 oC

0.3 K

Minimum sheath length 500 mm

Calibration Fee

561

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

1324.0 oC

0.6 K

By Co-C fixed point cell Minimum sheath length 700 mm

Calibration Fee

562

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

0 ?C to 660 ?C

0.2 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

563

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

> 660 ?C to 1100 ?C

0.3 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

564

11020 - 12311

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

> 1100 ?C to 1324 ?C

0.6 K

By ITS-90 Fixed Point Cells / Co-C fixed point cell (At least 3 points)

Calibration Fee

565

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

-40 ?C to 0 ?C

0.2 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

566

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

0 ?C to 250 ?C

0.1 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

567

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>250 ?C to 500 ?C

0.2 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

568

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>500 ?C to 700 ?C

0.5 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

569

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>700 ?C to 1100 ?C

1.0 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

570

11020 - 12312

Noble metal Thermocouples (Type S, Type R, Type B, Au/Pt)

>1100 ?C to 1324 ?C

1.5 K

By comparison with standard TC or SPRT in comparison furnace Minimum sheath length 450 mm

Calibration Fee

571

11020 - 12321

Base Metal Thermocouples

-40 ?C to 1100 ?C

0.1 - 1.5 K

By comparison with standard TC or SPRT in comparison furnace

Calibration Fee

572

11020 - 12411

Liquid-in-glass Thermometer

-40 ?C to 90 ?C

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

573

11020 - 12411

Liquid-in-glass Thermometer

> 90 ?C to 110 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

574

11020 - 12411

Liquid-in-glass Thermometer

> 110 ?C to 250 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

575

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

-80oC to -40oC (Resolution < 0.01oC)

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

576

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

- 40 ?C to 90 ?C

15 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

577

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

> 90 ?C to 250 ?C

20 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

578

11020 - 12711

Digital Thermometer (with probe) (resolution<0.2K)

> 250 ?C to 450 ?C

50 mK

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

579

11020 - 12712

Digital Thermometer (with probe) (resolution>0.2K)

-80oC to -40oC (Resolution > 0.1oC)

0.1 K

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

580

11020 - 12712

Digital Thermometer (with probe) (resolution>0.2K)

- 40 ?C to 450 ?C

0.1 K

By Comparison with SPRTs in stirred liquid bath

Calibration Fee

581

11020 - 22311

Supply of Thermocouple Type S with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

582

11020 - 22312

Supply of Thermocouple Type R with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

583

11020 - 22313

Supply of Thermocouple Type B with fixed point calibration

0 ?C to 1100 ?C

0.3 ?C

Supply of thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

584

11020 - 22314

Fabrication of Standard Thermocouples Type R, Type S, Type B with fixed point calibration (wires not included)

0 ?C to 1100 ?C

0.3 ?C

Electrical Annealing and re-fabrication thermocouple with calibration certificate by ITS-90 Fixed Point Cells.

Calibration Fee

585

11020 - 12861

Temperature Bath

-80oC to 25oC (Alcohol)

20 mK

By Comparison with SPRTs

Calibration Fee

586

11020 - 12861

Temperature Bath

>5 ?C to 50 ?C (Water)

15 mK

By Comparison with SPRTs

Calibration Fee

587

11020 - 12861

Temperature Bath

>40 ?C to 250 ?C (Oil)

30 mK

By Comparison with SPRTs

Calibration Fee

588

11020 - 12861

Temperature Bath

>300oC to 500oC (Salt)

50 mK

By Comparison with SPRTs

Calibration Fee

589

11020 - 12862

Dry Block and Furnace

- 40 ?C to 420 ?C

0.1 K

By Comparison with SPRTs

Calibration Fee

590

11020 - 12862

Dry Block and Furnace

>420oC to 660oC

0.1 K

By Comparison with SPRTs

Calibration Fee

591

11030 - 11121

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

1084.62oC

0.20 K

Comparison to Standard Fixed-point Blackbody using 0.65 ?m standard radiation thermometer

Calibration Fee

592

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

1 084.62oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

593

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

961.78oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

594

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

660.323oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

595

11030 - 11122

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

419.527oC

0.25 K

Comparison to Standard Fixed-point Blackbody using 0.9 ?m standard radiation thermometer

Calibration Fee

596

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

961.78oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

597

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

660.323oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

598

11030 - 11123

Fixed-point Blackbody (Primary and Secondary Fixed-point Black body)

419.527oC

0.30 K

Comparison to Standard Fixed-point Blackbody using 1.6 ?m standard radiation thermometer

Calibration Fee

599

11030 - 11411

Optical Pyrometer

156.5985 ?C to 1084.60 ?C

0.3 K to 0.5 K

By Comparison with Black Body Fixed Points

Calibration Fee

600

11030 - 11412

0.9 ?m radiation thermometer

960 ?C to 2000 ?C

0.39 K to 2.4 K

Direct realisation by measurement of spectral responsivity

Calibration Fee

601

11030 - 11413

0.65 ?m radiation thermometer

960 ?C to 2500 ?C

0.28 K to 1.6 K

Direct realisation by measurement of spectral responsivity

Calibration Fee

602

11030-12521

Variable Temperature Blackbody Radiation Sources

0 ?C to < 15 ?C

0.4 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 13 mm

Calibration Fee

603

11030-12525

Variable Temperature Blackbody Radiation Sources

15 ?C to 500 ?C

0.4 K to 2.1 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 13 mm

Calibration Fee

604

11030-12526

Variable Temperature Blackbody Radiation Sources

-20 ?C to < 15 ?C

0.4 K to 0.2 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 14 mm

Calibration Fee

605

11030-12527

Variable Temperature Blackbody Radiation Sources

15 ?C to 500 ?C

0.2 K to 2.1 K

Comparison with reference blackbodies through infrared thermometer spectral wavelength 8 mm to 14 mm

Calibration Fee

606

11030-12522

Variable Temperature Blackbody Radiation Sources

150 ?C to 1 100 ?C

0.7 K to 1.6 K

Comparison with 1.6 mm standard radiation thermometer

Calibration Fee

607

11030-12523

Variable Temperature Blackbody Radiation Sources

1 000 ?C to 2 500 ?C

1.6 K to 2.8 K

Comparison with 650 nm standard radiation thermometer

Calibration Fee

608

11030 - 12524

Blackbody sources for infrared ear thermometers

35 ? C to 42 ?C

55 mK

Comparison with standard liquid bath blackbodies through infrared ear thermometers

Calibration Fee

609

11030 - 12541

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

-5 ?C to 15 ?C

0.5 K to 1.5 K

By Comparison with PRT or Transfer Standard Radiation Thermometer

Calibration Fee

610

11030 - 12542

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

>15 ?C to < 1100 ?C

0.5 K to 5.0 K

By Comparison with Transfer Standard Radiation Thermometer or PRT

Calibration Fee

611

11030 - 12543

Radiation Thermometer/Infrared Thermometer/Thermal imaging Camera

1100 ?C to 2500 ?C

2 K to 5 K

By Comparison with Transfer Standard Radiation Thermometer

Calibration Fee

612

11030-12544

Infrared ear thermometer (Resolution ? 0.1 K)

35 ?C to 42 ?C

0.2 K

Comparison with PRT in liquid bath blackbody

Calibration Fee

613

11030-12545

Infrared ear thermometer (Resolution < 0.1 K)

35 ?C to 42 ?C

0.10 K

Comparison with PRT in liquid bath blackbody

Calibration Fee

614

11040 - 13111

Dew Point Hygrometer

- 75 ?C to -25 ?C

0.3 K

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

615

11040 - 13112

Dew Point Hygrometer

>- 25 ?C to +1 ?C

0.2 K

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

616

11040 - 13113

Dew Point Hygrometer

>1 ?C to 25 ?C

0.1 K

Comparison with two pressure Humidity Generator

Calibration Fee

617

11040 - 13114

Dew Point Hygrometer

>25 ?C to 30 ?C

0.2 K

Comparison with two pressure Humidity Generator

Calibration Fee

618

11040-13115

Dew-Point Hygrometer

>30 oC to 60 oC

0.1 oC

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

619

11040 - 13311

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

15 % to 50 %

0.5 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

620

11040 - 13312

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

> 50 % to 80 %

0.7 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

621

11040 - 13313

Relative Hygrometer(At chamber temperature : 15 oC to 60 oC )

> 80 % to 98 %

1.0 %

Comparison with two temperature two pressure Humidity Generator

Calibration Fee

622

11040 - 13411

Thermo-Hygrograph

Temperature 15 ?C to 50 ?C

(0.5 to 1) ?C

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

623

11040 - 13411

Thermo-Hygrometer

Humidity (35 to 95)%RH

(1.5 to 3) %RH

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

624

11040 - 13412

Data Logger (Air Type)

Temperature 15 ?C to 50 ?C

(0.5 to 1) ?C

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

625

11040 - 13412

Data Logger (Air Type)

Humidity (35 to 95)%RH

(1.5 to 3) %RH

Comparison with PRT and Chilled Mirror Hygrometer in Controlled Chamber

Calibration Fee

Top