บริการสอบเทียบของสถาบันมาตรวิทยาแห่งชาติ
Photometry Metrology
Item No. |
Code |
Description |
Range |
Accuracy |
Remark |
Service |
---|---|---|---|---|---|---|
1 |
13015-11011 |
Luminous intensity of an incandescent lamp |
10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K |
0.79% |
Substitution method against two national reference standard lamps |
Calibration Fee |
2 |
13016-11011 |
Luminous intensity of an incandescent lamp |
10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K |
0.93% |
Substitution method against two working standard lamps |
Calibration Fee |
3 |
13015-11021 |
Illuminance responsivity of a photo-detector head |
12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K |
0.70% |
Source-based measurements using national reference standard lamps applying the inverse-square law. |
Calibration Fee |
4 |
13016-11021 |
Illuminance responsivity of a photo-detector head |
12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K |
0.80% |
Source-based measurements using working standard lamps applying the inverse-square law. |
Calibration Fee |
5 |
13011-11021 |
Illuminance responsivity of an illuminance meter or a lux meter |
2 lx ? ?? ? 10 000 lx 2 000 K ? CCT ? 3 400 K |
Greater than or equal to 0.91% varied with illuminance level and UUC |
Detector-based comparison against reference photometers |
Calibration Fee |
6 |
13010-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.66% |
Detector-based method by using a gonio-photometer with a standard photometer |
Calibration Fee |
7 |
13015-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.80% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
8 |
13016-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.98% |
Substitution method inside a 2-meter integrating sphere against two working standard lamps |
Calibration Fee |
9 |
13015-11032 |
Total luminous flux of a non-directional LED lamp |
20 lm ? ?? ? 4 000 lm Full (non-directional), white |
0.79% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
10 |
13016-11032 |
Total luminous flux of a non-directional LED lamp |
20 lm ? ?? ? 4 000 lm Full (non-directional), white |
1.0% |
Substitution method inside a 2-meter integrating sphere against two working standard lamps |
Calibration Fee |
11 |
13015-11033 |
Total luminous flux of a linear fluorescent lamp |
800 lm ? ?? ? 4 000 lm 2 600 K ? CCT ? 6 200 K |
0.88% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
12 |
13011-11060 |
Luminance responsivity of a luminance meter |
10 cd/m2 ? L? ? 10 000 cd/m2 2 000 K ? CCT ? 3 400 K |
0.85% |
Detector-based comparison against reference photometers |
Calibration Fee |
13 |
13011-11060 |
Luminance responsivity of a luminance meter |
1 cd/m2 ? L? < 10 cd/m2 2 000 K ? CCT ? 3 400 K |
1.2% |
Detector-based comparison against reference photometers and a monitor detector |
Calibration Fee |
14 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
31.6% < %T ? 97.7% (0.01 ? OD < 0.5) |
200 nm to 2000 nm: Greater than or equal to 0.48% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
15 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
10.0% < %T ? 31.6% (0.5 ? OD < 1.0) |
200 nm to 2000: Greater than or equal to 0.57% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
16 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
1.00% < %T ? 10.00% (1.0 ? OD <2.0) |
200 nm to 2000: Greater than or equal to 0.61% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
17 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
0.01% < %T ?1.00% (2.0 ? OD <4.0) |
200 nm to 1800: Greater than or equal to 0.81% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
18 |
13025-14021 |
Diffuse spectral transmittance of a diffusely transmitting material |
380 nm to 780 nm |
For 20% haze plate, typical uncertainties in diffuse spectral transmittance are 0.13% to 0.18% throughout the spectrum |
Direct measurement by single-beam primary reference spectrophtometer |
Calibration Fee |
19 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
20 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
21 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
22 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
23 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
24 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
25 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
26 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
27 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
28 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
29 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
30 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
31 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
380 nm to 400 nm: 2.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
32 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
405 nm to 460 nm: 1.6% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
33 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
465 nm to 780 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
34 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
35 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
36 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
37 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
38 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
39 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
40 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
41 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
42 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
43 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
44 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
45 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
46 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
380 nm to 400 nm: 2.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
47 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
405 nm to 460 nm: 1.6% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
48 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
465 nm to 780 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
49 |
13025-14061 |
Regular spectral reflectance |
5 ? %R ? 100% 200 nm ? l ? 2 500 nm at 5 nm interval |
0.32% to 0.42% for the wavelength range from 200 nm to 885 nm 1.0% to 1.4% for the wavelength range from 890 nm to 2 500 nm *Typical values for a high reflective aluminum mirror. Uncertainties for other materials are likely higher. |
Absolute method by a double-subtractive double-beam spectrophotometer (Perkin-Elmer Lambda 1050) equipped with a variable-angle reflectance accessory (URA) |
Calibration Fee |
50 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
51 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
52 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
53 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
54 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
55 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
56 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
57 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
58 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
59 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
60 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
61 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
62 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
380 nm to 400 nm: 2.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
63 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
405 nm to 460 nm: 1.6% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
64 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
465 nm to 780 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
65 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
66 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
67 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
68 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
69 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
70 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
71 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
72 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
73 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
74 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
75 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
76 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
77 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
380 nm to 400 nm: 2.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
78 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
405 nm to 460 nm: 1.6% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
79 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
465 nm to 780 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
80 |
13025-14150 |
Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.) |
200 nm to 2 000 nm |
200 nm to 830 nm: 0.08 nm |
Measurement by reference spectrometer |
Calibration Fee |
81 |
13025-14150 |
Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.) |
200 nm to 2 000 nm |
830 nm to 2 000 nm: 0.50 nm |
Measurement by reference spectrometer |
Calibration Fee |
82 |
13025-14151 |
Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.) |
200 nm to 2 000 nm |
200 nm to 830 nm: 0.15 nm |
Measurement by reference spectrometer with integrating sphere accessory |
Calibration Fee |
83 |
13025-14151 |
Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.) |
200 nm to 2 000 nm |
830 nm to 2 000 nm: 0.70 nm |
Measurement by reference spectrometer with integrating sphere accessory |
Calibration Fee |
84 |
13025-16010 |
Surface color of general reflecting material, x, y, Y, u', v', L*, a*, b* |
Varied be parameter of interest, please contact a laboratory staff for full details. |
Varied by parameter of interest and color of UUC, please contact a laboratory staff for full details. |
Spectral diffuse reflectance measurement by single-beam substitution method against a reference color standard |
Calibration Fee |
85 |
13025-16030 |
Transmitted color of a transmitting material |
360 nm to 860 nm |
Values and uncertainties vary depending on sample's color. Contact the Colorimetry Laboratory for details |
Direct measurement by single-beam primary reference spectrophtometer |
Calibration Fee |
86 |
13025-16070 |
Haze and luminous transmittance of a diffusely transmitting material |
380 nm to 780 nm |
For 20% haze plate, typical uncertainty in transmission haze is 0.16%. |
Based on ASTM D1003 standard method |
Calibration Fee |
87 |
13025-16120 |
Whiteness and Tint |
380 nm to 780 nm W greater than 40 and less than 5Y-280 |
1.5 |
Spectral diffuse reflectance measurement by single-beam substitution method against a reference white diffuse reflectance standard |
Calibration Fee |
88 |
13038-12023 |
Absolute spectral irradiance responsivity of a scanning spectroradiometer with irradiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.5% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
89 |
13039-12023 |
Absolute spectral irradiance responsivity of a diode array spectroradiometer with irradiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.4% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
90 |
13038-12030 |
Relative spectral radiance responsivity of a scanning spectroradiometer with radiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.5% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
91 |
13039-12030 |
Relative spectral radiance responsivity of a diode array spectroradiometer with radiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.4% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
92 |
13034-12071 |
Solar irradiance responsivity of a solar power meter |
400 nm to 1 100 nm depending on UUC |
4.2% |
Comparison against a working standard spectroradiometer applying 300 W Xe arc lamp with airmass 1.5 Global filter |
Calibration Fee |
93 |
13034-12090 |
Intregrated irradiance responsivity of broadband radiometer |
315 nm to 400 nm (UVA ) |
4.4% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
94 |
13034-12091 |
Intregrated irradiance responsivity of broadband radiometer |
280 nm to 315 nm (UVB) |
6.3% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
95 |
13034-12092 |
Intregrated irradiance responsivity of broadband radiometer |
280 nm to 400 nm (UVA+UVB) |
6.0% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
96 |
13034-12099 |
Intregrated irradiance responsivity of broadband radiometer |
Arbitary range within 280 nm to 1 020 nm |
Varied by selected range |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
97 |
13034-12110 |
Intregrated irradiance responsivity of narrowband radiometer |
365 nm ? 10 nm |
4.4% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
98 |
13034-12113 |
Intregrated irradiance responsivity of narrowband radiometer |
254 nm ? 10 nm |
10% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
99 |
13034-12119 |
Intregrated irradiance responsivity of narrowband radiometer |
Arbitary range within 280 nm to 1 020 nm |
Varied by selected range |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
100 |
13038-12120 |
Responsivity of PAR radiometer |
400 nm to 700 nm |
2.0% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
101 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
250 nm ? l ? 320 nm, BW = 5 nm |
4.3% to 7.6% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
102 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
320 nm < l ? 400 nm, BW = 5 nm |
2.5% to 3.8% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
103 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
400 nm < l ? 1 100 nm, BW = 5 nm |
2.2% to 2.3% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
104 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
1 100 nm < l ? 1 800 nm, BW = 5 nm |
2.7% to 2.9% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
105 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
1 800 nm < l ? 2 300 nm, BW = 10 nm |
3.6% to 4.5% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
106 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
2 300 nm < l ? 2 500 nm, BW = 10 nm |
4.9% to 10% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
107 |
13037-13012 |
Spectral irradiance of a duterium lamp |
240 nm to 400 nm |
3.8% to 10% |
By scanning measurements against a spectral irradiance reference national standard lamp using a double subtractive monochromator |
Calibration Fee |
108 |
13034-13013 |
Measurement of absolute / relative spectral irradance of a light source |
250 nm to 1 020 nm |
Greater than or equal to 2.4% |
Measurement by a working standard diode array spectroradiometer |
Calibration Fee |
109 |
13038-15030 |
Correlated colour temperature response of a filter radiometer/Spectrometer |
3 000 K to 3 200 K |
20 K or 0.70% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
110 |
13035-15090 |
Measurement of integrated absolute irradance of a light source |
320 nm to 400 nm (UVA) |
4.2% |
Measurement by a working standard radiometer |
Calibration Fee |
111 |
13035-15091 |
Measurement of integrated absolute irradance of a light source |
260 nm to 320 nm (UVB) |
6.4% |
Measurement by a working standard radiometer |
Calibration Fee |
112 |
13035-15098 |
Measurement of integrated absolute irradance of a light source |
400 nm to 500 nm |
3.6% |
Measurement by a working standard radiometer |
Calibration Fee |
113 |
13032-12900 |
Measurement of linearity response of a photo-detector |
At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm |
200 nm to 400 nm: ? 0.1% 400 nm to 2 500 nm: ? 0.05% |
Flux superposition method |
Calibration Fee |
114 |
13032-12910 |
Measurement of uniformity of response of a photo-detector |
At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm |
201 nm to 400 nm: ? 0.2% 400 nm to 2 500 nm: ? 0.1% |
x-y linear scan |
Calibration Fee |
115 |
13045-11012 |
Luminous intensity of a single packaged LED |
0.1 cd to 100 cd |
2.3% to 3.4% varied with LED color |
Spectroradiometric method against two spectral irradiance standard lamps for CIE condition A and B |
Calibration Fee |
116 |
13045-11032 |
Luminous flux of a single-packaged LED |
0.1 lm to 200 lm |
1.8% to 2.3% varied with LED color |
Absolute integrating sphere method against two spectral irradiance standard lamps |
Calibration Fee |
117 |
13045-13051 |
Total spectral radiant flux of a tungsten lamp |
1.49 mW/nm ? ?e ? 35.8 mW/nm 360 nm ? l ? 830 nm BW = 1 nm |
3.1% to 6.2% |
Absolute integrating sphere method against spectral irradiance standard lamps |
Calibration Fee |
118 |
13041-15010 |
Distribution temperature of tungsten lamp |
2 600 K to 3 200 K |
20 K* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
119 |
13041-15021 |
Correlated colour temperature of general light source |
2 600 K to 6 600 K |
21 K* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
120 |
13041-15040 |
Chromaticity coordinate of general light source |
0 to 0.9 at approximately 1 nm bandwidth |
U(x)=0.001 7* U(y)=0.000 6* U(u)=0.001 0* U(v)=0.000 3* U(u')=0.001 0* U(v')=0.000 5* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
121 |
13045-15050 |
Chromaticity response of a colorimeter |
x=0.426 9 y=0.400 7 u'=0.245 5 v'=0.518 5 |
0.004 0 0.002 0 0.002 3 0.001 2 |
Source-based calibration against a standard lamp |
Calibration Fee |
122 |
13034-33013 |
Measurement of absolute / relative spectral irradance of a light source |
250 nm to 1020 nm |
3.4% to 9.8% |
Measurement by a working standard diode array spectroradiometer |
On-site Calibration |
123 |
13035-35090 |
Measurement of integrated absolute irradance of a light source |
320 nm to 400 nm (UVA) |
4.2% |
Measurement by a working standard radiometer |
On-site Calibration |
124 |
13035-35091 |
Measurement of integrated absolute irradance of a light source |
260 nm to 320 nm (UVB) |
6.4% |
Measurement by a working standard radiometer |
On-site Calibration |
125 |
13035-35098 |
Measurement of integrated absolute irradance of a light source |
400 nm to 500 nm |
3.6% |
Measurement by a working standard radiometer |
On-site Calibration |
126 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
400 nm to 1,100 nm for spectral match evaluation |
0.006* to 0.02* |
Measurements based on IEC 60904-9 |
On-site Calibration |
127 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
? 6,606 cm2 for designated test area of non-uniformity evaluation |
0.10% |
Measurements based on IEC 60904-9 |
On-site Calibration |
128 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
Data sampling time of DUT ? 2 ms for instability evaluation |
0.05% |
Measurements based on IEC 60904-9 |
On-site Calibration |
129 |
13015-11011 |
Luminous intensity of an incandescent lamp |
10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K |
0.79% |
Substitution method against two national reference standard lamps |
Calibration Fee |
130 |
13016-11011 |
Luminous intensity of an incandescent lamp |
10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K |
0.93% |
Substitution method against two working standard lamps |
Calibration Fee |
131 |
13015-11021 |
Illuminance responsivity of a photo-detector head |
12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K |
0.70% |
Source-based measurements using national reference standard lamps applying the inverse-square law. |
Calibration Fee |
132 |
13016-11021 |
Illuminance responsivity of a photo-detector head |
12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K |
0.80% |
Source-based measurements using working standard lamps applying the inverse-square law. |
Calibration Fee |
133 |
13011-11021 |
Illuminance responsivity of an illuminance meter or a lux meter |
2 lx ? ?? ? 10 000 lx 2 000 K ? CCT ? 3 400 K |
Greater than or equal to 0.91% varied with illuminance level and UUC |
Detector-based comparison against reference photometers |
Calibration Fee |
134 |
13010-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.66% |
Detector-based method by using a gonio-photometer with a standard photometer |
Calibration Fee |
135 |
13015-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.80% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
136 |
13016-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.98% |
Substitution method inside a 2-meter integrating sphere against two working standard lamps |
Calibration Fee |
137 |
13015-11032 |
Total luminous flux of a non-directional LED lamp |
20 lm ? ?? ? 4 000 lm Full (non-directional), white |
0.79% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
138 |
13016-11032 |
Total luminous flux of a non-directional LED lamp |
20 lm ? ?? ? 4 000 lm Full (non-directional), white |
1.0% |
Substitution method inside a 2-meter integrating sphere against two working standard lamps |
Calibration Fee |
139 |
13015-11033 |
Total luminous flux of a linear fluorescent lamp |
800 lm ? ?? ? 4 000 lm 2 600 K ? CCT ? 6 200 K |
0.88% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
140 |
13011-11060 |
Luminance responsivity of a luminance meter |
10 cd/m2 ? L? ? 10 000 cd/m2 2 000 K ? CCT ? 3 400 K |
0.85% |
Detector-based comparison against reference photometers |
Calibration Fee |
141 |
13011-11060 |
Luminance responsivity of a luminance meter |
1 cd/m2 ? L? < 10 cd/m2 2 000 K ? CCT ? 3 400 K |
1.2% |
Detector-based comparison against reference photometers and a monitor detector |
Calibration Fee |
142 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
31.6% < %T ? 97.7% (0.01 ? OD < 0.5) |
200 nm to 2000 nm: Greater than or equal to 0.48% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
143 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
10.0% < %T ? 31.6% (0.5 ? OD < 1.0) |
200 nm to 2000: Greater than or equal to 0.57% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
144 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
1.00% < %T ? 10.00% (1.0 ? OD <2.0) |
200 nm to 2000: Greater than or equal to 0.61% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
145 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
0.01% < %T ?1.00% (2.0 ? OD <4.0) |
200 nm to 1800: Greater than or equal to 0.81% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
146 |
13025-14021 |
Diffuse spectral transmittance of a diffusely transmitting material |
380 nm to 780 nm |
For 20% haze plate, typical uncertainties in diffuse spectral transmittance are 0.13% to 0.18% throughout the spectrum |
Direct measurement by single-beam primary reference spectrophtometer |
Calibration Fee |
147 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
148 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
149 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
150 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
151 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
152 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
153 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
154 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
155 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
156 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
157 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
158 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
159 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
380 nm to 400 nm: 2.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
160 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
405 nm to 460 nm: 1.6% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
161 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
465 nm to 780 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
162 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
163 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
164 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
165 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
166 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
167 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
168 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
169 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
170 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
171 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
172 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
173 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
174 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
380 nm to 400 nm: 2.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
175 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
405 nm to 460 nm: 1.6% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
176 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
465 nm to 780 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
177 |
13025-14061 |
Regular spectral reflectance |
5 ? %R ? 100% 200 nm ? l ? 2 500 nm at 5 nm interval |
0.32% to 0.42% for the wavelength range from 200 nm to 885 nm 1.0% to 1.4% for the wavelength range from 890 nm to 2 500 nm *Typical values for a high reflective aluminum mirror. Uncertainties for other materials are likely higher. |
Absolute method by a double-subtractive double-beam spectrophotometer (Perkin-Elmer Lambda 1050) equipped with a variable-angle reflectance accessory (URA) |
Calibration Fee |
178 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
179 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
180 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
181 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
182 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
183 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
184 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
185 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
186 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
187 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
188 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
189 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
190 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
380 nm to 400 nm: 2.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
191 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
405 nm to 460 nm: 1.6% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
192 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
465 nm to 780 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
193 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
194 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
195 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
196 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
197 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
198 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
199 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
200 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
201 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
202 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
203 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
204 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
205 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
380 nm to 400 nm: 2.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
206 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
405 nm to 460 nm: 1.6% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
207 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
465 nm to 780 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
208 |
13025-14150 |
Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.) |
200 nm to 2 000 nm |
200 nm to 830 nm: 0.08 nm |
Measurement by reference spectrometer |
Calibration Fee |
209 |
13025-14150 |
Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.) |
200 nm to 2 000 nm |
830 nm to 2 000 nm: 0.50 nm |
Measurement by reference spectrometer |
Calibration Fee |
210 |
13025-14151 |
Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.) |
200 nm to 2 000 nm |
200 nm to 830 nm: 0.15 nm |
Measurement by reference spectrometer with integrating sphere accessory |
Calibration Fee |
211 |
13025-14151 |
Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.) |
200 nm to 2 000 nm |
830 nm to 2 000 nm: 0.70 nm |
Measurement by reference spectrometer with integrating sphere accessory |
Calibration Fee |
212 |
13025-16010 |
Surface color of general reflecting material, x, y, Y, u', v', L*, a*, b* |
Varied be parameter of interest, please contact a laboratory staff for full details. |
Varied by parameter of interest and color of UUC, please contact a laboratory staff for full details. |
Spectral diffuse reflectance measurement by single-beam substitution method against a reference color standard |
Calibration Fee |
213 |
13025-16030 |
Transmitted color of a transmitting material |
360 nm to 860 nm |
Values and uncertainties vary depending on sample's color. Contact the Colorimetry Laboratory for details |
Direct measurement by single-beam primary reference spectrophtometer |
Calibration Fee |
214 |
13025-16070 |
Haze and luminous transmittance of a diffusely transmitting material |
380 nm to 780 nm |
For 20% haze plate, typical uncertainty in transmission haze is 0.16%. |
Based on ASTM D1003 standard method |
Calibration Fee |
215 |
13025-16120 |
Whiteness and Tint |
380 nm to 780 nm W greater than 40 and less than 5Y-280 |
1.5 |
Spectral diffuse reflectance measurement by single-beam substitution method against a reference white diffuse reflectance standard |
Calibration Fee |
216 |
13038-12023 |
Absolute spectral irradiance responsivity of a scanning spectroradiometer with irradiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.5% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
217 |
13039-12023 |
Absolute spectral irradiance responsivity of a diode array spectroradiometer with irradiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.4% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
218 |
13038-12030 |
Relative spectral radiance responsivity of a scanning spectroradiometer with radiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.5% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
219 |
13039-12030 |
Relative spectral radiance responsivity of a diode array spectroradiometer with radiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.4% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
220 |
13034-12071 |
Solar irradiance responsivity of a solar power meter |
400 nm to 1 100 nm depending on UUC |
4.2% |
Comparison against a working standard spectroradiometer applying 300 W Xe arc lamp with airmass 1.5 Global filter |
Calibration Fee |
221 |
13034-12090 |
Intregrated irradiance responsivity of broadband radiometer |
315 nm to 400 nm (UVA ) |
4.4% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
222 |
13034-12091 |
Intregrated irradiance responsivity of broadband radiometer |
280 nm to 315 nm (UVB) |
6.3% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
223 |
13034-12092 |
Intregrated irradiance responsivity of broadband radiometer |
280 nm to 400 nm (UVA+UVB) |
6.0% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
224 |
13034-12099 |
Intregrated irradiance responsivity of broadband radiometer |
Arbitary range within 280 nm to 1 020 nm |
Varied by selected range |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
225 |
13034-12110 |
Intregrated irradiance responsivity of narrowband radiometer |
365 nm ? 10 nm |
4.4% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
226 |
13034-12113 |
Intregrated irradiance responsivity of narrowband radiometer |
254 nm ? 10 nm |
10% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
227 |
13034-12119 |
Intregrated irradiance responsivity of narrowband radiometer |
Arbitary range within 280 nm to 1 020 nm |
Varied by selected range |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
228 |
13038-12120 |
Responsivity of PAR radiometer |
400 nm to 700 nm |
2.0% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
229 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
250 nm ? l ? 320 nm, BW = 5 nm |
4.3% to 7.6% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
230 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
320 nm < l ? 400 nm, BW = 5 nm |
2.5% to 3.8% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
231 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
400 nm < l ? 1 100 nm, BW = 5 nm |
2.2% to 2.3% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
232 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
1 100 nm < l ? 1 800 nm, BW = 5 nm |
2.7% to 2.9% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
233 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
1 800 nm < l ? 2 300 nm, BW = 10 nm |
3.6% to 4.5% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
234 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
2 300 nm < l ? 2 500 nm, BW = 10 nm |
4.9% to 10% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
235 |
13037-13012 |
Spectral irradiance of a duterium lamp |
240 nm to 400 nm |
3.8% to 10% |
By scanning measurements against a spectral irradiance reference national standard lamp using a double subtractive monochromator |
Calibration Fee |
236 |
13034-13013 |
Measurement of absolute / relative spectral irradance of a light source |
250 nm to 1 020 nm |
Greater than or equal to 2.4% |
Measurement by a working standard diode array spectroradiometer |
Calibration Fee |
237 |
13038-15030 |
Correlated colour temperature response of a filter radiometer/Spectrometer |
3 000 K to 3 200 K |
20 K or 0.70% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
238 |
13035-15090 |
Measurement of integrated absolute irradance of a light source |
320 nm to 400 nm (UVA) |
4.2% |
Measurement by a working standard radiometer |
Calibration Fee |
239 |
13035-15091 |
Measurement of integrated absolute irradance of a light source |
260 nm to 320 nm (UVB) |
6.4% |
Measurement by a working standard radiometer |
Calibration Fee |
240 |
13035-15098 |
Measurement of integrated absolute irradance of a light source |
400 nm to 500 nm |
3.6% |
Measurement by a working standard radiometer |
Calibration Fee |
241 |
13032-12900 |
Measurement of linearity response of a photo-detector |
At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm |
200 nm to 400 nm: ? 0.1% 400 nm to 2 500 nm: ? 0.05% |
Flux superposition method |
Calibration Fee |
242 |
13032-12910 |
Measurement of uniformity of response of a photo-detector |
At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm |
201 nm to 400 nm: ? 0.2% 400 nm to 2 500 nm: ? 0.1% |
x-y linear scan |
Calibration Fee |
243 |
13045-11012 |
Luminous intensity of a single packaged LED |
0.1 cd to 100 cd |
2.3% to 3.4% varied with LED color |
Spectroradiometric method against two spectral irradiance standard lamps for CIE condition A and B |
Calibration Fee |
244 |
13045-11032 |
Luminous flux of a single-packaged LED |
0.1 lm to 200 lm |
1.8% to 2.3% varied with LED color |
Absolute integrating sphere method against two spectral irradiance standard lamps |
Calibration Fee |
245 |
13045-13051 |
Total spectral radiant flux of a tungsten lamp |
1.49 mW/nm ? ?e ? 35.8 mW/nm 360 nm ? l ? 830 nm BW = 1 nm |
3.1% to 6.2% |
Absolute integrating sphere method against spectral irradiance standard lamps |
Calibration Fee |
246 |
13041-15010 |
Distribution temperature of tungsten lamp |
2 600 K to 3 200 K |
20 K* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
247 |
13041-15021 |
Correlated colour temperature of general light source |
2 600 K to 6 600 K |
21 K* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
248 |
13041-15040 |
Chromaticity coordinate of general light source |
0 to 0.9 at approximately 1 nm bandwidth |
U(x)=0.001 7* U(y)=0.000 6* U(u)=0.001 0* U(v)=0.000 3* U(u')=0.001 0* U(v')=0.000 5* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
249 |
13045-15050 |
Chromaticity response of a colorimeter |
x=0.426 9 y=0.400 7 u'=0.245 5 v'=0.518 5 |
0.004 0 0.002 0 0.002 3 0.001 2 |
Source-based calibration against a standard lamp |
Calibration Fee |
250 |
13034-33013 |
Measurement of absolute / relative spectral irradance of a light source |
250 nm to 1020 nm |
3.4% to 9.8% |
Measurement by a working standard diode array spectroradiometer |
On-site Calibration |
251 |
13035-35090 |
Measurement of integrated absolute irradance of a light source |
320 nm to 400 nm (UVA) |
4.2% |
Measurement by a working standard radiometer |
On-site Calibration |
252 |
13035-35091 |
Measurement of integrated absolute irradance of a light source |
260 nm to 320 nm (UVB) |
6.4% |
Measurement by a working standard radiometer |
On-site Calibration |
253 |
13035-35098 |
Measurement of integrated absolute irradance of a light source |
400 nm to 500 nm |
3.6% |
Measurement by a working standard radiometer |
On-site Calibration |
254 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
400 nm to 1,100 nm for spectral match evaluation |
0.006* to 0.02* |
Measurements based on IEC 60904-9 |
On-site Calibration |
255 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
? 6,606 cm2 for designated test area of non-uniformity evaluation |
0.10% |
Measurements based on IEC 60904-9 |
On-site Calibration |
256 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
Data sampling time of DUT ? 2 ms for instability evaluation |
0.05% |
Measurements based on IEC 60904-9 |
On-site Calibration |
257 |
13015-11011 |
Luminous intensity of an incandescent lamp |
10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K |
0.79% |
Substitution method against two national reference standard lamps |
Calibration Fee |
258 |
13016-11011 |
Luminous intensity of an incandescent lamp |
10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K |
0.93% |
Substitution method against two working standard lamps |
Calibration Fee |
259 |
13015-11021 |
Illuminance responsivity of a photo-detector head |
12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K |
0.70% |
Source-based measurements using national reference standard lamps applying the inverse-square law. |
Calibration Fee |
260 |
13016-11021 |
Illuminance responsivity of a photo-detector head |
12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K |
0.80% |
Source-based measurements using working standard lamps applying the inverse-square law. |
Calibration Fee |
261 |
13011-11021 |
Illuminance responsivity of an illuminance meter or a lux meter |
2 lx ? ?? ? 10 000 lx 2 000 K ? CCT ? 3 400 K |
Greater than or equal to 0.91% varied with illuminance level and UUC |
Detector-based comparison against reference photometers |
Calibration Fee |
262 |
13010-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.66% |
Detector-based method by using a gonio-photometer with a standard photometer |
Calibration Fee |
263 |
13015-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.80% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
264 |
13016-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.98% |
Substitution method inside a 2-meter integrating sphere against two working standard lamps |
Calibration Fee |
265 |
13015-11032 |
Total luminous flux of a non-directional LED lamp |
20 lm ? ?? ? 4 000 lm Full (non-directional), white |
0.79% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
266 |
13016-11032 |
Total luminous flux of a non-directional LED lamp |
20 lm ? ?? ? 4 000 lm Full (non-directional), white |
1.0% |
Substitution method inside a 2-meter integrating sphere against two working standard lamps |
Calibration Fee |
267 |
13015-11033 |
Total luminous flux of a linear fluorescent lamp |
800 lm ? ?? ? 4 000 lm 2 600 K ? CCT ? 6 200 K |
0.88% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
268 |
13011-11060 |
Luminance responsivity of a luminance meter |
10 cd/m2 ? L? ? 10 000 cd/m2 2 000 K ? CCT ? 3 400 K |
0.85% |
Detector-based comparison against reference photometers |
Calibration Fee |
269 |
13011-11060 |
Luminance responsivity of a luminance meter |
1 cd/m2 ? L? < 10 cd/m2 2 000 K ? CCT ? 3 400 K |
1.2% |
Detector-based comparison against reference photometers and a monitor detector |
Calibration Fee |
270 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
31.6% < %T ? 97.7% (0.01 ? OD < 0.5) |
200 nm to 2000 nm: Greater than or equal to 0.48% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
271 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
10.0% < %T ? 31.6% (0.5 ? OD < 1.0) |
200 nm to 2000: Greater than or equal to 0.57% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
272 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
1.00% < %T ? 10.00% (1.0 ? OD <2.0) |
200 nm to 2000: Greater than or equal to 0.61% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
273 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
0.01% < %T ?1.00% (2.0 ? OD <4.0) |
200 nm to 1800: Greater than or equal to 0.81% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
274 |
13025-14021 |
Diffuse spectral transmittance of a diffusely transmitting material |
380 nm to 780 nm |
For 20% haze plate, typical uncertainties in diffuse spectral transmittance are 0.13% to 0.18% throughout the spectrum |
Direct measurement by single-beam primary reference spectrophtometer |
Calibration Fee |
275 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
276 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
277 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
278 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
279 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
280 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
281 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
282 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
283 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
284 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
285 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
286 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
287 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
380 nm to 400 nm: 2.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
288 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
405 nm to 460 nm: 1.6% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
289 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
465 nm to 780 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
290 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
291 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
292 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
293 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
294 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
295 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
296 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
297 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
298 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
299 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
300 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
301 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
302 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
380 nm to 400 nm: 2.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
303 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
405 nm to 460 nm: 1.6% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
304 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
465 nm to 780 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
305 |
13025-14061 |
Regular spectral reflectance |
5 ? %R ? 100% 200 nm ? l ? 2 500 nm at 5 nm interval |
0.32% to 0.42% for the wavelength range from 200 nm to 885 nm 1.0% to 1.4% for the wavelength range from 890 nm to 2 500 nm *Typical values for a high reflective aluminum mirror. Uncertainties for other materials are likely higher. |
Absolute method by a double-subtractive double-beam spectrophotometer (Perkin-Elmer Lambda 1050) equipped with a variable-angle reflectance accessory (URA) |
Calibration Fee |
306 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
307 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
308 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
309 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
310 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
311 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
312 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
313 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
314 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
315 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
316 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
317 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
318 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
380 nm to 400 nm: 2.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
319 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
405 nm to 460 nm: 1.6% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
320 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
465 nm to 780 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
321 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
322 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
323 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
324 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
325 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
326 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
327 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
328 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
329 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
330 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
331 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
332 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
333 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
380 nm to 400 nm: 2.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
334 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
405 nm to 460 nm: 1.6% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
335 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
465 nm to 780 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
336 |
13025-14150 |
Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.) |
200 nm to 2 000 nm |
200 nm to 830 nm: 0.08 nm |
Measurement by reference spectrometer |
Calibration Fee |
337 |
13025-14150 |
Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.) |
200 nm to 2 000 nm |
830 nm to 2 000 nm: 0.50 nm |
Measurement by reference spectrometer |
Calibration Fee |
338 |
13025-14151 |
Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.) |
200 nm to 2 000 nm |
200 nm to 830 nm: 0.15 nm |
Measurement by reference spectrometer with integrating sphere accessory |
Calibration Fee |
339 |
13025-14151 |
Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.) |
200 nm to 2 000 nm |
830 nm to 2 000 nm: 0.70 nm |
Measurement by reference spectrometer with integrating sphere accessory |
Calibration Fee |
340 |
13025-16010 |
Surface color of general reflecting material, x, y, Y, u', v', L*, a*, b* |
Varied be parameter of interest, please contact a laboratory staff for full details. |
Varied by parameter of interest and color of UUC, please contact a laboratory staff for full details. |
Spectral diffuse reflectance measurement by single-beam substitution method against a reference color standard |
Calibration Fee |
341 |
13025-16030 |
Transmitted color of a transmitting material |
360 nm to 860 nm |
Values and uncertainties vary depending on sample's color. Contact the Colorimetry Laboratory for details |
Direct measurement by single-beam primary reference spectrophtometer |
Calibration Fee |
342 |
13025-16070 |
Haze and luminous transmittance of a diffusely transmitting material |
380 nm to 780 nm |
For 20% haze plate, typical uncertainty in transmission haze is 0.16%. |
Based on ASTM D1003 standard method |
Calibration Fee |
343 |
13025-16120 |
Whiteness and Tint |
380 nm to 780 nm W greater than 40 and less than 5Y-280 |
1.5 |
Spectral diffuse reflectance measurement by single-beam substitution method against a reference white diffuse reflectance standard |
Calibration Fee |
344 |
13038-12023 |
Absolute spectral irradiance responsivity of a scanning spectroradiometer with irradiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.5% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
345 |
13039-12023 |
Absolute spectral irradiance responsivity of a diode array spectroradiometer with irradiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.4% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
346 |
13038-12030 |
Relative spectral radiance responsivity of a scanning spectroradiometer with radiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.5% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
347 |
13039-12030 |
Relative spectral radiance responsivity of a diode array spectroradiometer with radiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.4% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
348 |
13034-12071 |
Solar irradiance responsivity of a solar power meter |
400 nm to 1 100 nm depending on UUC |
4.2% |
Comparison against a working standard spectroradiometer applying 300 W Xe arc lamp with airmass 1.5 Global filter |
Calibration Fee |
349 |
13034-12090 |
Intregrated irradiance responsivity of broadband radiometer |
315 nm to 400 nm (UVA ) |
4.4% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
350 |
13034-12091 |
Intregrated irradiance responsivity of broadband radiometer |
280 nm to 315 nm (UVB) |
6.3% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
351 |
13034-12092 |
Intregrated irradiance responsivity of broadband radiometer |
280 nm to 400 nm (UVA+UVB) |
6.0% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
352 |
13034-12099 |
Intregrated irradiance responsivity of broadband radiometer |
Arbitary range within 280 nm to 1 020 nm |
Varied by selected range |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
353 |
13034-12110 |
Intregrated irradiance responsivity of narrowband radiometer |
365 nm ? 10 nm |
4.4% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
354 |
13034-12113 |
Intregrated irradiance responsivity of narrowband radiometer |
254 nm ? 10 nm |
10% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
355 |
13034-12119 |
Intregrated irradiance responsivity of narrowband radiometer |
Arbitary range within 280 nm to 1 020 nm |
Varied by selected range |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
356 |
13038-12120 |
Responsivity of PAR radiometer |
400 nm to 700 nm |
2.0% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
357 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
250 nm ? l ? 320 nm, BW = 5 nm |
4.3% to 7.6% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
358 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
320 nm < l ? 400 nm, BW = 5 nm |
2.5% to 3.8% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
359 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
400 nm < l ? 1 100 nm, BW = 5 nm |
2.2% to 2.3% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
360 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
1 100 nm < l ? 1 800 nm, BW = 5 nm |
2.7% to 2.9% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
361 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
1 800 nm < l ? 2 300 nm, BW = 10 nm |
3.6% to 4.5% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
362 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
2 300 nm < l ? 2 500 nm, BW = 10 nm |
4.9% to 10% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
363 |
13037-13012 |
Spectral irradiance of a duterium lamp |
240 nm to 400 nm |
3.8% to 10% |
By scanning measurements against a spectral irradiance reference national standard lamp using a double subtractive monochromator |
Calibration Fee |
364 |
13034-13013 |
Measurement of absolute / relative spectral irradance of a light source |
250 nm to 1 020 nm |
Greater than or equal to 2.4% |
Measurement by a working standard diode array spectroradiometer |
Calibration Fee |
365 |
13038-15030 |
Correlated colour temperature response of a filter radiometer/Spectrometer |
3 000 K to 3 200 K |
20 K or 0.70% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
366 |
13035-15090 |
Measurement of integrated absolute irradance of a light source |
320 nm to 400 nm (UVA) |
4.2% |
Measurement by a working standard radiometer |
Calibration Fee |
367 |
13035-15091 |
Measurement of integrated absolute irradance of a light source |
260 nm to 320 nm (UVB) |
6.4% |
Measurement by a working standard radiometer |
Calibration Fee |
368 |
13035-15098 |
Measurement of integrated absolute irradance of a light source |
400 nm to 500 nm |
3.6% |
Measurement by a working standard radiometer |
Calibration Fee |
369 |
13032-12900 |
Measurement of linearity response of a photo-detector |
At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm |
200 nm to 400 nm: ? 0.1% 400 nm to 2 500 nm: ? 0.05% |
Flux superposition method |
Calibration Fee |
370 |
13032-12910 |
Measurement of uniformity of response of a photo-detector |
At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm |
201 nm to 400 nm: ? 0.2% 400 nm to 2 500 nm: ? 0.1% |
x-y linear scan |
Calibration Fee |
371 |
13045-11012 |
Luminous intensity of a single packaged LED |
0.1 cd to 100 cd |
2.3% to 3.4% varied with LED color |
Spectroradiometric method against two spectral irradiance standard lamps for CIE condition A and B |
Calibration Fee |
372 |
13045-11032 |
Luminous flux of a single-packaged LED |
0.1 lm to 200 lm |
1.8% to 2.3% varied with LED color |
Absolute integrating sphere method against two spectral irradiance standard lamps |
Calibration Fee |
373 |
13045-13051 |
Total spectral radiant flux of a tungsten lamp |
1.49 mW/nm ? ?e ? 35.8 mW/nm 360 nm ? l ? 830 nm BW = 1 nm |
3.1% to 6.2% |
Absolute integrating sphere method against spectral irradiance standard lamps |
Calibration Fee |
374 |
13041-15010 |
Distribution temperature of tungsten lamp |
2 600 K to 3 200 K |
20 K* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
375 |
13041-15021 |
Correlated colour temperature of general light source |
2 600 K to 6 600 K |
21 K* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
376 |
13041-15040 |
Chromaticity coordinate of general light source |
0 to 0.9 at approximately 1 nm bandwidth |
U(x)=0.001 7* U(y)=0.000 6* U(u)=0.001 0* U(v)=0.000 3* U(u')=0.001 0* U(v')=0.000 5* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
377 |
13045-15050 |
Chromaticity response of a colorimeter |
x=0.426 9 y=0.400 7 u'=0.245 5 v'=0.518 5 |
0.004 0 0.002 0 0.002 3 0.001 2 |
Source-based calibration against a standard lamp |
Calibration Fee |
378 |
13034-33013 |
Measurement of absolute / relative spectral irradance of a light source |
250 nm to 1020 nm |
3.4% to 9.8% |
Measurement by a working standard diode array spectroradiometer |
On-site Calibration |
379 |
13035-35090 |
Measurement of integrated absolute irradance of a light source |
320 nm to 400 nm (UVA) |
4.2% |
Measurement by a working standard radiometer |
On-site Calibration |
380 |
13035-35091 |
Measurement of integrated absolute irradance of a light source |
260 nm to 320 nm (UVB) |
6.4% |
Measurement by a working standard radiometer |
On-site Calibration |
381 |
13035-35098 |
Measurement of integrated absolute irradance of a light source |
400 nm to 500 nm |
3.6% |
Measurement by a working standard radiometer |
On-site Calibration |
382 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
400 nm to 1,100 nm for spectral match evaluation |
0.006* to 0.02* |
Measurements based on IEC 60904-9 |
On-site Calibration |
383 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
? 6,606 cm2 for designated test area of non-uniformity evaluation |
0.10% |
Measurements based on IEC 60904-9 |
On-site Calibration |
384 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
Data sampling time of DUT ? 2 ms for instability evaluation |
0.05% |
Measurements based on IEC 60904-9 |
On-site Calibration |
385 |
13015-11011 |
Luminous intensity of an incandescent lamp |
10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K |
0.79% |
Substitution method against two national reference standard lamps |
Calibration Fee |
386 |
13016-11011 |
Luminous intensity of an incandescent lamp |
10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K |
0.93% |
Substitution method against two working standard lamps |
Calibration Fee |
387 |
13015-11021 |
Illuminance responsivity of a photo-detector head |
12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K |
0.70% |
Source-based measurements using national reference standard lamps applying the inverse-square law. |
Calibration Fee |
388 |
13016-11021 |
Illuminance responsivity of a photo-detector head |
12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K |
0.80% |
Source-based measurements using working standard lamps applying the inverse-square law. |
Calibration Fee |
389 |
13011-11021 |
Illuminance responsivity of an illuminance meter or a lux meter |
2 lx ? ?? ? 10 000 lx 2 000 K ? CCT ? 3 400 K |
Greater than or equal to 0.91% varied with illuminance level and UUC |
Detector-based comparison against reference photometers |
Calibration Fee |
390 |
13010-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.66% |
Detector-based method by using a gonio-photometer with a standard photometer |
Calibration Fee |
391 |
13015-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.80% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
392 |
13016-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.98% |
Substitution method inside a 2-meter integrating sphere against two working standard lamps |
Calibration Fee |
393 |
13015-11032 |
Total luminous flux of a non-directional LED lamp |
20 lm ? ?? ? 4 000 lm Full (non-directional), white |
0.79% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
394 |
13016-11032 |
Total luminous flux of a non-directional LED lamp |
20 lm ? ?? ? 4 000 lm Full (non-directional), white |
1.0% |
Substitution method inside a 2-meter integrating sphere against two working standard lamps |
Calibration Fee |
395 |
13015-11033 |
Total luminous flux of a linear fluorescent lamp |
800 lm ? ?? ? 4 000 lm 2 600 K ? CCT ? 6 200 K |
0.88% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
396 |
13011-11060 |
Luminance responsivity of a luminance meter |
10 cd/m2 ? L? ? 10 000 cd/m2 2 000 K ? CCT ? 3 400 K |
0.85% |
Detector-based comparison against reference photometers |
Calibration Fee |
397 |
13011-11060 |
Luminance responsivity of a luminance meter |
1 cd/m2 ? L? < 10 cd/m2 2 000 K ? CCT ? 3 400 K |
1.2% |
Detector-based comparison against reference photometers and a monitor detector |
Calibration Fee |
398 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
31.6% < %T ? 97.7% (0.01 ? OD < 0.5) |
200 nm to 2000 nm: Greater than or equal to 0.48% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
399 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
10.0% < %T ? 31.6% (0.5 ? OD < 1.0) |
200 nm to 2000: Greater than or equal to 0.57% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
400 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
1.00% < %T ? 10.00% (1.0 ? OD <2.0) |
200 nm to 2000: Greater than or equal to 0.61% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
401 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
0.01% < %T ?1.00% (2.0 ? OD <4.0) |
200 nm to 1800: Greater than or equal to 0.81% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
402 |
13025-14021 |
Diffuse spectral transmittance of a diffusely transmitting material |
380 nm to 780 nm |
For 20% haze plate, typical uncertainties in diffuse spectral transmittance are 0.13% to 0.18% throughout the spectrum |
Direct measurement by single-beam primary reference spectrophtometer |
Calibration Fee |
403 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
404 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
405 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
406 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
407 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
408 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
409 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
410 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
411 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
412 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
413 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
414 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
415 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
380 nm to 400 nm: 2.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
416 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
405 nm to 460 nm: 1.6% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
417 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
465 nm to 780 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
418 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
419 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
420 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
421 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
422 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
423 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
424 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
425 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
426 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
427 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
428 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
429 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
430 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
380 nm to 400 nm: 2.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
431 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
405 nm to 460 nm: 1.6% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
432 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
465 nm to 780 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
433 |
13025-14061 |
Regular spectral reflectance |
5 ? %R ? 100% 200 nm ? l ? 2 500 nm at 5 nm interval |
0.32% to 0.42% for the wavelength range from 200 nm to 885 nm 1.0% to 1.4% for the wavelength range from 890 nm to 2 500 nm *Typical values for a high reflective aluminum mirror. Uncertainties for other materials are likely higher. |
Absolute method by a double-subtractive double-beam spectrophotometer (Perkin-Elmer Lambda 1050) equipped with a variable-angle reflectance accessory (URA) |
Calibration Fee |
434 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
435 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
436 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
437 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
438 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
439 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
440 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
441 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
442 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
443 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
444 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
445 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
446 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
380 nm to 400 nm: 2.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
447 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
405 nm to 460 nm: 1.6% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
448 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
465 nm to 780 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
449 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
450 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
451 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
452 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
453 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
454 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
455 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
456 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
457 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
458 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
459 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
460 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
461 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
380 nm to 400 nm: 2.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
462 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
405 nm to 460 nm: 1.6% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
463 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
465 nm to 780 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
464 |
13025-14150 |
Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.) |
200 nm to 2 000 nm |
200 nm to 830 nm: 0.08 nm |
Measurement by reference spectrometer |
Calibration Fee |
465 |
13025-14150 |
Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.) |
200 nm to 2 000 nm |
830 nm to 2 000 nm: 0.50 nm |
Measurement by reference spectrometer |
Calibration Fee |
466 |
13025-14151 |
Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.) |
200 nm to 2 000 nm |
200 nm to 830 nm: 0.15 nm |
Measurement by reference spectrometer with integrating sphere accessory |
Calibration Fee |
467 |
13025-14151 |
Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.) |
200 nm to 2 000 nm |
830 nm to 2 000 nm: 0.70 nm |
Measurement by reference spectrometer with integrating sphere accessory |
Calibration Fee |
468 |
13025-16010 |
Surface color of general reflecting material, x, y, Y, u', v', L*, a*, b* |
Varied be parameter of interest, please contact a laboratory staff for full details. |
Varied by parameter of interest and color of UUC, please contact a laboratory staff for full details. |
Spectral diffuse reflectance measurement by single-beam substitution method against a reference color standard |
Calibration Fee |
469 |
13025-16030 |
Transmitted color of a transmitting material |
360 nm to 860 nm |
Values and uncertainties vary depending on sample's color. Contact the Colorimetry Laboratory for details |
Direct measurement by single-beam primary reference spectrophtometer |
Calibration Fee |
470 |
13025-16070 |
Haze and luminous transmittance of a diffusely transmitting material |
380 nm to 780 nm |
For 20% haze plate, typical uncertainty in transmission haze is 0.16%. |
Based on ASTM D1003 standard method |
Calibration Fee |
471 |
13025-16120 |
Whiteness and Tint |
380 nm to 780 nm W greater than 40 and less than 5Y-280 |
1.5 |
Spectral diffuse reflectance measurement by single-beam substitution method against a reference white diffuse reflectance standard |
Calibration Fee |
472 |
13038-12023 |
Absolute spectral irradiance responsivity of a scanning spectroradiometer with irradiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.5% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
473 |
13039-12023 |
Absolute spectral irradiance responsivity of a diode array spectroradiometer with irradiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.4% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
474 |
13038-12030 |
Relative spectral radiance responsivity of a scanning spectroradiometer with radiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.5% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
475 |
13039-12030 |
Relative spectral radiance responsivity of a diode array spectroradiometer with radiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.4% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
476 |
13034-12071 |
Solar irradiance responsivity of a solar power meter |
400 nm to 1 100 nm depending on UUC |
4.2% |
Comparison against a working standard spectroradiometer applying 300 W Xe arc lamp with airmass 1.5 Global filter |
Calibration Fee |
477 |
13034-12090 |
Intregrated irradiance responsivity of broadband radiometer |
315 nm to 400 nm (UVA ) |
4.4% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
478 |
13034-12091 |
Intregrated irradiance responsivity of broadband radiometer |
280 nm to 315 nm (UVB) |
6.3% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
479 |
13034-12092 |
Intregrated irradiance responsivity of broadband radiometer |
280 nm to 400 nm (UVA+UVB) |
6.0% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
480 |
13034-12099 |
Intregrated irradiance responsivity of broadband radiometer |
Arbitary range within 280 nm to 1 020 nm |
Varied by selected range |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
481 |
13034-12110 |
Intregrated irradiance responsivity of narrowband radiometer |
365 nm ? 10 nm |
4.4% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
482 |
13034-12113 |
Intregrated irradiance responsivity of narrowband radiometer |
254 nm ? 10 nm |
10% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
483 |
13034-12119 |
Intregrated irradiance responsivity of narrowband radiometer |
Arbitary range within 280 nm to 1 020 nm |
Varied by selected range |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
484 |
13038-12120 |
Responsivity of PAR radiometer |
400 nm to 700 nm |
2.0% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
485 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
250 nm ? l ? 320 nm, BW = 5 nm |
4.3% to 7.6% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
486 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
320 nm < l ? 400 nm, BW = 5 nm |
2.5% to 3.8% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
487 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
400 nm < l ? 1 100 nm, BW = 5 nm |
2.2% to 2.3% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
488 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
1 100 nm < l ? 1 800 nm, BW = 5 nm |
2.7% to 2.9% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
489 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
1 800 nm < l ? 2 300 nm, BW = 10 nm |
3.6% to 4.5% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
490 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
2 300 nm < l ? 2 500 nm, BW = 10 nm |
4.9% to 10% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
491 |
13037-13012 |
Spectral irradiance of a duterium lamp |
240 nm to 400 nm |
3.8% to 10% |
By scanning measurements against a spectral irradiance reference national standard lamp using a double subtractive monochromator |
Calibration Fee |
492 |
13034-13013 |
Measurement of absolute / relative spectral irradance of a light source |
250 nm to 1 020 nm |
Greater than or equal to 2.4% |
Measurement by a working standard diode array spectroradiometer |
Calibration Fee |
493 |
13038-15030 |
Correlated colour temperature response of a filter radiometer/Spectrometer |
3 000 K to 3 200 K |
20 K or 0.70% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
494 |
13035-15090 |
Measurement of integrated absolute irradance of a light source |
320 nm to 400 nm (UVA) |
4.2% |
Measurement by a working standard radiometer |
Calibration Fee |
495 |
13035-15091 |
Measurement of integrated absolute irradance of a light source |
260 nm to 320 nm (UVB) |
6.4% |
Measurement by a working standard radiometer |
Calibration Fee |
496 |
13035-15098 |
Measurement of integrated absolute irradance of a light source |
400 nm to 500 nm |
3.6% |
Measurement by a working standard radiometer |
Calibration Fee |
497 |
13032-12900 |
Measurement of linearity response of a photo-detector |
At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm |
200 nm to 400 nm: ? 0.1% 400 nm to 2 500 nm: ? 0.05% |
Flux superposition method |
Calibration Fee |
498 |
13032-12910 |
Measurement of uniformity of response of a photo-detector |
At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm |
201 nm to 400 nm: ? 0.2% 400 nm to 2 500 nm: ? 0.1% |
x-y linear scan |
Calibration Fee |
499 |
13045-11012 |
Luminous intensity of a single packaged LED |
0.1 cd to 100 cd |
2.3% to 3.4% varied with LED color |
Spectroradiometric method against two spectral irradiance standard lamps for CIE condition A and B |
Calibration Fee |
500 |
13045-11032 |
Luminous flux of a single-packaged LED |
0.1 lm to 200 lm |
1.8% to 2.3% varied with LED color |
Absolute integrating sphere method against two spectral irradiance standard lamps |
Calibration Fee |
501 |
13045-13051 |
Total spectral radiant flux of a tungsten lamp |
1.49 mW/nm ? ?e ? 35.8 mW/nm 360 nm ? l ? 830 nm BW = 1 nm |
3.1% to 6.2% |
Absolute integrating sphere method against spectral irradiance standard lamps |
Calibration Fee |
502 |
13041-15010 |
Distribution temperature of tungsten lamp |
2 600 K to 3 200 K |
20 K* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
503 |
13041-15021 |
Correlated colour temperature of general light source |
2 600 K to 6 600 K |
21 K* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
504 |
13041-15040 |
Chromaticity coordinate of general light source |
0 to 0.9 at approximately 1 nm bandwidth |
U(x)=0.001 7* U(y)=0.000 6* U(u)=0.001 0* U(v)=0.000 3* U(u')=0.001 0* U(v')=0.000 5* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
505 |
13045-15050 |
Chromaticity response of a colorimeter |
x=0.426 9 y=0.400 7 u'=0.245 5 v'=0.518 5 |
0.004 0 0.002 0 0.002 3 0.001 2 |
Source-based calibration against a standard lamp |
Calibration Fee |
506 |
13034-33013 |
Measurement of absolute / relative spectral irradance of a light source |
250 nm to 1020 nm |
3.4% to 9.8% |
Measurement by a working standard diode array spectroradiometer |
On-site Calibration |
507 |
13035-35090 |
Measurement of integrated absolute irradance of a light source |
320 nm to 400 nm (UVA) |
4.2% |
Measurement by a working standard radiometer |
On-site Calibration |
508 |
13035-35091 |
Measurement of integrated absolute irradance of a light source |
260 nm to 320 nm (UVB) |
6.4% |
Measurement by a working standard radiometer |
On-site Calibration |
509 |
13035-35098 |
Measurement of integrated absolute irradance of a light source |
400 nm to 500 nm |
3.6% |
Measurement by a working standard radiometer |
On-site Calibration |
510 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
400 nm to 1,100 nm for spectral match evaluation |
0.006* to 0.02* |
Measurements based on IEC 60904-9 |
On-site Calibration |
511 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
? 6,606 cm2 for designated test area of non-uniformity evaluation |
0.10% |
Measurements based on IEC 60904-9 |
On-site Calibration |
512 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
Data sampling time of DUT ? 2 ms for instability evaluation |
0.05% |
Measurements based on IEC 60904-9 |
On-site Calibration |
513 |
13015-11011 |
Luminous intensity of an incandescent lamp |
10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K |
0.79% |
Substitution method against two national reference standard lamps |
Calibration Fee |
514 |
13016-11011 |
Luminous intensity of an incandescent lamp |
10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K |
0.93% |
Substitution method against two working standard lamps |
Calibration Fee |
515 |
13015-11021 |
Illuminance responsivity of a photo-detector head |
12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K |
0.70% |
Source-based measurements using national reference standard lamps applying the inverse-square law. |
Calibration Fee |
516 |
13016-11021 |
Illuminance responsivity of a photo-detector head |
12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K |
0.80% |
Source-based measurements using working standard lamps applying the inverse-square law. |
Calibration Fee |
517 |
13011-11021 |
Illuminance responsivity of an illuminance meter or a lux meter |
2 lx ? ?? ? 10 000 lx 2 000 K ? CCT ? 3 400 K |
Greater than or equal to 0.91% varied with illuminance level and UUC |
Detector-based comparison against reference photometers |
Calibration Fee |
518 |
13010-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.66% |
Detector-based method by using a gonio-photometer with a standard photometer |
Calibration Fee |
519 |
13015-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.80% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
520 |
13016-11031 |
Total luminous flux of an incandescent lamp |
5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K |
0.98% |
Substitution method inside a 2-meter integrating sphere against two working standard lamps |
Calibration Fee |
521 |
13015-11032 |
Total luminous flux of a non-directional LED lamp |
20 lm ? ?? ? 4 000 lm Full (non-directional), white |
0.79% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
522 |
13016-11032 |
Total luminous flux of a non-directional LED lamp |
20 lm ? ?? ? 4 000 lm Full (non-directional), white |
1.0% |
Substitution method inside a 2-meter integrating sphere against two working standard lamps |
Calibration Fee |
523 |
13015-11033 |
Total luminous flux of a linear fluorescent lamp |
800 lm ? ?? ? 4 000 lm 2 600 K ? CCT ? 6 200 K |
0.88% |
Substitution method inside a 2-meter integrating sphere against two national reference standard lamps |
Calibration Fee |
524 |
13011-11060 |
Luminance responsivity of a luminance meter |
10 cd/m2 ? L? ? 10 000 cd/m2 2 000 K ? CCT ? 3 400 K |
0.85% |
Detector-based comparison against reference photometers |
Calibration Fee |
525 |
13011-11060 |
Luminance responsivity of a luminance meter |
1 cd/m2 ? L? < 10 cd/m2 2 000 K ? CCT ? 3 400 K |
1.2% |
Detector-based comparison against reference photometers and a monitor detector |
Calibration Fee |
526 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
31.6% < %T ? 97.7% (0.01 ? OD < 0.5) |
200 nm to 2000 nm: Greater than or equal to 0.48% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
527 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
10.0% < %T ? 31.6% (0.5 ? OD < 1.0) |
200 nm to 2000: Greater than or equal to 0.57% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
528 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
1.00% < %T ? 10.00% (1.0 ? OD <2.0) |
200 nm to 2000: Greater than or equal to 0.61% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
529 |
13025-14011 |
Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm. |
0.01% < %T ?1.00% (2.0 ? OD <4.0) |
200 nm to 1800: Greater than or equal to 0.81% |
By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
530 |
13025-14021 |
Diffuse spectral transmittance of a diffusely transmitting material |
380 nm to 780 nm |
For 20% haze plate, typical uncertainties in diffuse spectral transmittance are 0.13% to 0.18% throughout the spectrum |
Direct measurement by single-beam primary reference spectrophtometer |
Calibration Fee |
531 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
532 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
533 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
534 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
535 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
536 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
537 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
538 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
539 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
540 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
541 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
542 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
543 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
380 nm to 400 nm: 2.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
544 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
405 nm to 460 nm: 1.6% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
545 |
13025-14051 |
Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
465 nm to 780 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
546 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
547 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
548 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
549 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
550 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
551 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
552 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
553 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
554 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
555 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
556 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
557 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
10% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
558 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
380 nm to 400 nm: 2.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
559 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
405 nm to 460 nm: 1.6% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
560 |
13026-14051 |
Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
5% ? %R ? 10% |
465 nm to 780 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
561 |
13025-14061 |
Regular spectral reflectance |
5 ? %R ? 100% 200 nm ? l ? 2 500 nm at 5 nm interval |
0.32% to 0.42% for the wavelength range from 200 nm to 885 nm 1.0% to 1.4% for the wavelength range from 890 nm to 2 500 nm *Typical values for a high reflective aluminum mirror. Uncertainties for other materials are likely higher. |
Absolute method by a double-subtractive double-beam spectrophotometer (Perkin-Elmer Lambda 1050) equipped with a variable-angle reflectance accessory (URA) |
Calibration Fee |
562 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
563 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
564 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
565 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
566 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
567 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
568 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
569 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
570 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
571 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
572 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
573 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
574 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
380 nm to 400 nm: 2.1% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
575 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
405 nm to 460 nm: 1.6% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
576 |
13025-14071 |
Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
465 nm to 780 nm: 1.5% |
By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
577 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
578 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
579 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
75% < %R ? 100% |
465 nm to 780 nm: 0.86% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
580 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
380 nm to 400 nm: 1.7% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
581 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
405 nm to 460 nm: 1.2% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
582 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
40% < %R ? 75% |
465 nm to 780 nm:0.84% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
583 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
380 nm to 400 nm: 1.8% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
584 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
405 nm to 460 nm: 1.3% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
585 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
20% < %R ? 40% |
465 nm to 780 nm: 0.90% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
586 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
380 nm to 400 nm: 1.9% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
587 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
405 nm to 460 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
588 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
7.5% < %R ? 20% |
465 nm to 780 nm: 1.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
589 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
380 nm to 400 nm: 2.1% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
590 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
405 nm to 460 nm: 1.6% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
591 |
13026-14071 |
Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm. |
2.5% ? %R ? 7.5% |
465 nm to 780 nm: 1.5% |
By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC. |
Calibration Fee |
592 |
13025-14150 |
Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.) |
200 nm to 2 000 nm |
200 nm to 830 nm: 0.08 nm |
Measurement by reference spectrometer |
Calibration Fee |
593 |
13025-14150 |
Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.) |
200 nm to 2 000 nm |
830 nm to 2 000 nm: 0.50 nm |
Measurement by reference spectrometer |
Calibration Fee |
594 |
13025-14151 |
Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.) |
200 nm to 2 000 nm |
200 nm to 830 nm: 0.15 nm |
Measurement by reference spectrometer with integrating sphere accessory |
Calibration Fee |
595 |
13025-14151 |
Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.) |
200 nm to 2 000 nm |
830 nm to 2 000 nm: 0.70 nm |
Measurement by reference spectrometer with integrating sphere accessory |
Calibration Fee |
596 |
13025-16010 |
Surface color of general reflecting material, x, y, Y, u', v', L*, a*, b* |
Varied be parameter of interest, please contact a laboratory staff for full details. |
Varied by parameter of interest and color of UUC, please contact a laboratory staff for full details. |
Spectral diffuse reflectance measurement by single-beam substitution method against a reference color standard |
Calibration Fee |
597 |
13025-16030 |
Transmitted color of a transmitting material |
360 nm to 860 nm |
Values and uncertainties vary depending on sample's color. Contact the Colorimetry Laboratory for details |
Direct measurement by single-beam primary reference spectrophtometer |
Calibration Fee |
598 |
13025-16070 |
Haze and luminous transmittance of a diffusely transmitting material |
380 nm to 780 nm |
For 20% haze plate, typical uncertainty in transmission haze is 0.16%. |
Based on ASTM D1003 standard method |
Calibration Fee |
599 |
13025-16120 |
Whiteness and Tint |
380 nm to 780 nm W greater than 40 and less than 5Y-280 |
1.5 |
Spectral diffuse reflectance measurement by single-beam substitution method against a reference white diffuse reflectance standard |
Calibration Fee |
600 |
13038-12023 |
Absolute spectral irradiance responsivity of a scanning spectroradiometer with irradiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.5% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
601 |
13039-12023 |
Absolute spectral irradiance responsivity of a diode array spectroradiometer with irradiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.4% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
602 |
13038-12030 |
Relative spectral radiance responsivity of a scanning spectroradiometer with radiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.5% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
603 |
13039-12030 |
Relative spectral radiance responsivity of a diode array spectroradiometer with radiance probe |
250 nm to 2 500 nm |
Greater than or equal to 2.4% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
604 |
13034-12071 |
Solar irradiance responsivity of a solar power meter |
400 nm to 1 100 nm depending on UUC |
4.2% |
Comparison against a working standard spectroradiometer applying 300 W Xe arc lamp with airmass 1.5 Global filter |
Calibration Fee |
605 |
13034-12090 |
Intregrated irradiance responsivity of broadband radiometer |
315 nm to 400 nm (UVA ) |
4.4% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
606 |
13034-12091 |
Intregrated irradiance responsivity of broadband radiometer |
280 nm to 315 nm (UVB) |
6.3% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
607 |
13034-12092 |
Intregrated irradiance responsivity of broadband radiometer |
280 nm to 400 nm (UVA+UVB) |
6.0% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
608 |
13034-12099 |
Intregrated irradiance responsivity of broadband radiometer |
Arbitary range within 280 nm to 1 020 nm |
Varied by selected range |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
609 |
13034-12110 |
Intregrated irradiance responsivity of narrowband radiometer |
365 nm ? 10 nm |
4.4% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
610 |
13034-12113 |
Intregrated irradiance responsivity of narrowband radiometer |
254 nm ? 10 nm |
10% |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
611 |
13034-12119 |
Intregrated irradiance responsivity of narrowband radiometer |
Arbitary range within 280 nm to 1 020 nm |
Varied by selected range |
Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source |
Calibration Fee |
612 |
13038-12120 |
Responsivity of PAR radiometer |
400 nm to 700 nm |
2.0% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
613 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
250 nm ? l ? 320 nm, BW = 5 nm |
4.3% to 7.6% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
614 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
320 nm < l ? 400 nm, BW = 5 nm |
2.5% to 3.8% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
615 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
400 nm < l ? 1 100 nm, BW = 5 nm |
2.2% to 2.3% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
616 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
1 100 nm < l ? 1 800 nm, BW = 5 nm |
2.7% to 2.9% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
617 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
1 800 nm < l ? 2 300 nm, BW = 10 nm |
3.6% to 4.5% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
618 |
13037-13011 |
Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type) |
2 300 nm < l ? 2 500 nm, BW = 10 nm |
4.9% to 10% |
By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator |
Calibration Fee |
619 |
13037-13012 |
Spectral irradiance of a duterium lamp |
240 nm to 400 nm |
3.8% to 10% |
By scanning measurements against a spectral irradiance reference national standard lamp using a double subtractive monochromator |
Calibration Fee |
620 |
13034-13013 |
Measurement of absolute / relative spectral irradance of a light source |
250 nm to 1 020 nm |
Greater than or equal to 2.4% |
Measurement by a working standard diode array spectroradiometer |
Calibration Fee |
621 |
13038-15030 |
Correlated colour temperature response of a filter radiometer/Spectrometer |
3 000 K to 3 200 K |
20 K or 0.70% |
Direct measurement against a set of two FEL working standard lamps |
Calibration Fee |
622 |
13035-15090 |
Measurement of integrated absolute irradance of a light source |
320 nm to 400 nm (UVA) |
4.2% |
Measurement by a working standard radiometer |
Calibration Fee |
623 |
13035-15091 |
Measurement of integrated absolute irradance of a light source |
260 nm to 320 nm (UVB) |
6.4% |
Measurement by a working standard radiometer |
Calibration Fee |
624 |
13035-15098 |
Measurement of integrated absolute irradance of a light source |
400 nm to 500 nm |
3.6% |
Measurement by a working standard radiometer |
Calibration Fee |
625 |
13032-12900 |
Measurement of linearity response of a photo-detector |
At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm |
200 nm to 400 nm: ? 0.1% 400 nm to 2 500 nm: ? 0.05% |
Flux superposition method |
Calibration Fee |
626 |
13032-12910 |
Measurement of uniformity of response of a photo-detector |
At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm |
201 nm to 400 nm: ? 0.2% 400 nm to 2 500 nm: ? 0.1% |
x-y linear scan |
Calibration Fee |
627 |
13045-11012 |
Luminous intensity of a single packaged LED |
0.1 cd to 100 cd |
2.3% to 3.4% varied with LED color |
Spectroradiometric method against two spectral irradiance standard lamps for CIE condition A and B |
Calibration Fee |
628 |
13045-11032 |
Luminous flux of a single-packaged LED |
0.1 lm to 200 lm |
1.8% to 2.3% varied with LED color |
Absolute integrating sphere method against two spectral irradiance standard lamps |
Calibration Fee |
629 |
13045-13051 |
Total spectral radiant flux of a tungsten lamp |
1.49 mW/nm ? ?e ? 35.8 mW/nm 360 nm ? l ? 830 nm BW = 1 nm |
3.1% to 6.2% |
Absolute integrating sphere method against spectral irradiance standard lamps |
Calibration Fee |
630 |
13041-15010 |
Distribution temperature of tungsten lamp |
2 600 K to 3 200 K |
20 K* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
631 |
13041-15021 |
Correlated colour temperature of general light source |
2 600 K to 6 600 K |
21 K* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
632 |
13041-15040 |
Chromaticity coordinate of general light source |
0 to 0.9 at approximately 1 nm bandwidth |
U(x)=0.001 7* U(y)=0.000 6* U(u)=0.001 0* U(v)=0.000 3* U(u')=0.001 0* U(v')=0.000 5* varied by UUC |
Measurement by a working standard spectroradiometer |
Calibration Fee |
633 |
13045-15050 |
Chromaticity response of a colorimeter |
x=0.426 9 y=0.400 7 u'=0.245 5 v'=0.518 5 |
0.004 0 0.002 0 0.002 3 0.001 2 |
Source-based calibration against a standard lamp |
Calibration Fee |
634 |
13034-33013 |
Measurement of absolute / relative spectral irradance of a light source |
250 nm to 1020 nm |
3.4% to 9.8% |
Measurement by a working standard diode array spectroradiometer |
On-site Calibration |
635 |
13035-35090 |
Measurement of integrated absolute irradance of a light source |
320 nm to 400 nm (UVA) |
4.2% |
Measurement by a working standard radiometer |
On-site Calibration |
636 |
13035-35091 |
Measurement of integrated absolute irradance of a light source |
260 nm to 320 nm (UVB) |
6.4% |
Measurement by a working standard radiometer |
On-site Calibration |
637 |
13035-35098 |
Measurement of integrated absolute irradance of a light source |
400 nm to 500 nm |
3.6% |
Measurement by a working standard radiometer |
On-site Calibration |
638 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
400 nm to 1,100 nm for spectral match evaluation |
0.006* to 0.02* |
Measurements based on IEC 60904-9 |
On-site Calibration |
639 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
? 6,606 cm2 for designated test area of non-uniformity evaluation |
0.10% |
Measurements based on IEC 60904-9 |
On-site Calibration |
640 |
13034-33014 |
Solar simulator performance testing according to IEC 60904-9 |
Data sampling time of DUT ? 2 ms for instability evaluation |
0.05% |
Measurements based on IEC 60904-9 |
On-site Calibration |