บริการสอบเทียบของสถาบันมาตรวิทยาแห่งชาติ

Photometry Metrology

* แหล่งข้อมูล : รายการบริการสอบเทียบสถาบันมาตรวิทยาแห่งชาติ

Item No.

Code

Description

Range

Accuracy
/Uncertainty

Remark

Service

1

13015-11011

Luminous intensity of an incandescent lamp

10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K

0.79%

Substitution method against two national reference standard lamps

Calibration Fee

2

13016-11011

Luminous intensity of an incandescent lamp

10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K

0.93%

Substitution method against two working standard lamps

Calibration Fee

3

13015-11021

Illuminance responsivity of a photo-detector head

12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K

0.70%

Source-based measurements using national reference standard lamps applying the inverse-square law.

Calibration Fee

4

13016-11021

Illuminance responsivity of a photo-detector head

12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K

0.80%

Source-based measurements using working standard lamps applying the inverse-square law.

Calibration Fee

5

13011-11021

Illuminance responsivity of an illuminance meter or a lux meter

2 lx ? ?? ? 10 000 lx 2 000 K ? CCT ? 3 400 K

Greater than or equal to 0.91% varied with illuminance level and UUC

Detector-based comparison against reference photometers

Calibration Fee

6

13010-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.66%

Detector-based method by using a gonio-photometer with a standard photometer

Calibration Fee

7

13015-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.80%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

8

13016-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.98%

Substitution method inside a 2-meter integrating sphere against two working standard lamps

Calibration Fee

9

13015-11032

Total luminous flux of a non-directional LED lamp

20 lm ? ?? ? 4 000 lm Full (non-directional), white

0.79%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

10

13016-11032

Total luminous flux of a non-directional LED lamp

20 lm ? ?? ? 4 000 lm Full (non-directional), white

1.0%

Substitution method inside a 2-meter integrating sphere against two working standard lamps

Calibration Fee

11

13015-11033

Total luminous flux of a linear fluorescent lamp

800 lm ? ?? ? 4 000 lm 2 600 K ? CCT ? 6 200 K

0.88%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

12

13011-11060

Luminance responsivity of a luminance meter

10 cd/m2 ? L? ? 10 000 cd/m2 2 000 K ? CCT ? 3 400 K

0.85%

Detector-based comparison against reference photometers

Calibration Fee

13

13011-11060

Luminance responsivity of a luminance meter

1 cd/m2 ? L? < 10 cd/m2 2 000 K ? CCT ? 3 400 K

1.2%

Detector-based comparison against reference photometers and a monitor detector

Calibration Fee

14

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

31.6% < %T ? 97.7% (0.01 ? OD < 0.5)

200 nm to 2000 nm: Greater than or equal to 0.48%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

15

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

10.0% < %T ? 31.6% (0.5 ? OD < 1.0)

200 nm to 2000: Greater than or equal to 0.57%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

16

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

1.00% < %T ? 10.00% (1.0 ? OD <2.0)

200 nm to 2000: Greater than or equal to 0.61%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

17

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

0.01% < %T ?1.00% (2.0 ? OD <4.0)

200 nm to 1800: Greater than or equal to 0.81%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

18

13025-14021

Diffuse spectral transmittance of a diffusely transmitting material

380 nm to 780 nm

For 20% haze plate, typical uncertainties in diffuse spectral transmittance are 0.13% to 0.18% throughout the spectrum

Direct measurement by single-beam primary reference spectrophtometer

Calibration Fee

19

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

20

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

21

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

22

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

23

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

24

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

25

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

26

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

27

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

28

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

29

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

30

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

31

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

380 nm to 400 nm: 2.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

32

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

405 nm to 460 nm: 1.6%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

33

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

465 nm to 780 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

34

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

35

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

36

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

37

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

38

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

39

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

40

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

41

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

42

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

43

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

44

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

45

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

46

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

380 nm to 400 nm: 2.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

47

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

405 nm to 460 nm: 1.6%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

48

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

465 nm to 780 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

49

13025-14061

Regular spectral reflectance

5 ? %R ? 100% 200 nm ? l ? 2 500 nm at 5 nm interval

0.32% to 0.42% for the wavelength range from 200 nm to 885 nm 1.0% to 1.4% for the wavelength range from 890 nm to 2 500 nm *Typical values for a high reflective aluminum mirror. Uncertainties for other materials are likely higher.

Absolute method by a double-subtractive double-beam spectrophotometer (Perkin-Elmer Lambda 1050) equipped with a variable-angle reflectance accessory (URA)

Calibration Fee

50

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

51

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

52

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

53

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

54

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

55

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

56

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

57

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

58

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

59

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

60

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

61

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

62

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

380 nm to 400 nm: 2.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

63

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

405 nm to 460 nm: 1.6%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

64

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

465 nm to 780 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

65

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

66

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

67

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

68

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

69

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

70

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

71

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

72

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

73

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

74

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

75

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

76

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

77

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

380 nm to 400 nm: 2.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

78

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

405 nm to 460 nm: 1.6%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

79

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

465 nm to 780 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

80

13025-14150

Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.)

200 nm to 2 000 nm

200 nm to 830 nm: 0.08 nm

Measurement by reference spectrometer

Calibration Fee

81

13025-14150

Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.)

200 nm to 2 000 nm

830 nm to 2 000 nm: 0.50 nm

Measurement by reference spectrometer

Calibration Fee

82

13025-14151

Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.)

200 nm to 2 000 nm

200 nm to 830 nm: 0.15 nm

Measurement by reference spectrometer with integrating sphere accessory

Calibration Fee

83

13025-14151

Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.)

200 nm to 2 000 nm

830 nm to 2 000 nm: 0.70 nm

Measurement by reference spectrometer with integrating sphere accessory

Calibration Fee

84

13025-16010

Surface color of general reflecting material, x, y, Y, u', v', L*, a*, b*

Varied be parameter of interest, please contact a laboratory staff for full details.

Varied by parameter of interest and color of UUC, please contact a laboratory staff for full details.

Spectral diffuse reflectance measurement by single-beam substitution method against a reference color standard

Calibration Fee

85

13025-16030

Transmitted color of a transmitting material

360 nm to 860 nm

Values and uncertainties vary depending on sample's color. Contact the Colorimetry Laboratory for details

Direct measurement by single-beam primary reference spectrophtometer

Calibration Fee

86

13025-16070

Haze and luminous transmittance of a diffusely transmitting material

380 nm to 780 nm

For 20% haze plate, typical uncertainty in transmission haze is 0.16%.

Based on ASTM D1003 standard method

Calibration Fee

87

13025-16120

Whiteness and Tint

380 nm to 780 nm W greater than 40 and less than 5Y-280

1.5

Spectral diffuse reflectance measurement by single-beam substitution method against a reference white diffuse reflectance standard

Calibration Fee

88

13038-12023

Absolute spectral irradiance responsivity of a scanning spectroradiometer with irradiance probe

250 nm to 2 500 nm

Greater than or equal to 2.5%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

89

13039-12023

Absolute spectral irradiance responsivity of a diode array spectroradiometer with irradiance probe

250 nm to 2 500 nm

Greater than or equal to 2.4%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

90

13038-12030

Relative spectral radiance responsivity of a scanning spectroradiometer with radiance probe

250 nm to 2 500 nm

Greater than or equal to 2.5%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

91

13039-12030

Relative spectral radiance responsivity of a diode array spectroradiometer with radiance probe

250 nm to 2 500 nm

Greater than or equal to 2.4%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

92

13034-12071

Solar irradiance responsivity of a solar power meter

400 nm to 1 100 nm depending on UUC

4.2%

Comparison against a working standard spectroradiometer applying 300 W Xe arc lamp with airmass 1.5 Global filter

Calibration Fee

93

13034-12090

Intregrated irradiance responsivity of broadband radiometer

315 nm to 400 nm (UVA )

4.4%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

94

13034-12091

Intregrated irradiance responsivity of broadband radiometer

280 nm to 315 nm (UVB)

6.3%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

95

13034-12092

Intregrated irradiance responsivity of broadband radiometer

280 nm to 400 nm (UVA+UVB)

6.0%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

96

13034-12099

Intregrated irradiance responsivity of broadband radiometer

Arbitary range within 280 nm to 1 020 nm

Varied by selected range

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

97

13034-12110

Intregrated irradiance responsivity of narrowband radiometer

365 nm ? 10 nm

4.4%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

98

13034-12113

Intregrated irradiance responsivity of narrowband radiometer

254 nm ? 10 nm

10%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

99

13034-12119

Intregrated irradiance responsivity of narrowband radiometer

Arbitary range within 280 nm to 1 020 nm

Varied by selected range

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

100

13038-12120

Responsivity of PAR radiometer

400 nm to 700 nm

2.0%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

101

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

250 nm ? l ? 320 nm, BW = 5 nm

4.3% to 7.6%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

102

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

320 nm < l ? 400 nm, BW = 5 nm

2.5% to 3.8%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

103

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

400 nm < l ? 1 100 nm, BW = 5 nm

2.2% to 2.3%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

104

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

1 100 nm < l ? 1 800 nm, BW = 5 nm

2.7% to 2.9%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

105

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

1 800 nm < l ? 2 300 nm, BW = 10 nm

3.6% to 4.5%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

106

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

2 300 nm < l ? 2 500 nm, BW = 10 nm

4.9% to 10%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

107

13037-13012

Spectral irradiance of a duterium lamp

240 nm to 400 nm

3.8% to 10%

By scanning measurements against a spectral irradiance reference national standard lamp using a double subtractive monochromator

Calibration Fee

108

13034-13013

Measurement of absolute / relative spectral irradance of a light source

250 nm to 1 020 nm

Greater than or equal to 2.4%

Measurement by a working standard diode array spectroradiometer

Calibration Fee

109

13038-15030

Correlated colour temperature response of a filter radiometer/Spectrometer

3 000 K to 3 200 K

20 K or 0.70%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

110

13035-15090

Measurement of integrated absolute irradance of a light source

320 nm to 400 nm (UVA)

4.2%

Measurement by a working standard radiometer

Calibration Fee

111

13035-15091

Measurement of integrated absolute irradance of a light source

260 nm to 320 nm (UVB)

6.4%

Measurement by a working standard radiometer

Calibration Fee

112

13035-15098

Measurement of integrated absolute irradance of a light source

400 nm to 500 nm

3.6%

Measurement by a working standard radiometer

Calibration Fee

113

13032-12900

Measurement of linearity response of a photo-detector

At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm

200 nm to 400 nm: ? 0.1% 400 nm to 2 500 nm: ? 0.05%

Flux superposition method

Calibration Fee

114

13032-12910

Measurement of uniformity of response of a photo-detector

At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm

201 nm to 400 nm: ? 0.2% 400 nm to 2 500 nm: ? 0.1%

x-y linear scan

Calibration Fee

115

13045-11012

Luminous intensity of a single packaged LED

0.1 cd to 100 cd

2.3% to 3.4% varied with LED color

Spectroradiometric method against two spectral irradiance standard lamps for CIE condition A and B

Calibration Fee

116

13045-11032

Luminous flux of a single-packaged LED

0.1 lm to 200 lm

1.8% to 2.3% varied with LED color

Absolute integrating sphere method against two spectral irradiance standard lamps

Calibration Fee

117

13045-13051

Total spectral radiant flux of a tungsten lamp

1.49 mW/nm ? ?e ? 35.8 mW/nm 360 nm ? l ? 830 nm BW = 1 nm

3.1% to 6.2%

Absolute integrating sphere method against spectral irradiance standard lamps

Calibration Fee

118

13041-15010

Distribution temperature of tungsten lamp

2 600 K to 3 200 K

20 K* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

119

13041-15021

Correlated colour temperature of general light source

2 600 K to 6 600 K

21 K* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

120

13041-15040

Chromaticity coordinate of general light source

0 to 0.9 at approximately 1 nm bandwidth

U(x)=0.001 7* U(y)=0.000 6* U(u)=0.001 0* U(v)=0.000 3* U(u')=0.001 0* U(v')=0.000 5* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

121

13045-15050

Chromaticity response of a colorimeter

x=0.426 9 y=0.400 7 u'=0.245 5 v'=0.518 5

0.004 0 0.002 0 0.002 3 0.001 2

Source-based calibration against a standard lamp

Calibration Fee

122

13034-33013

Measurement of absolute / relative spectral irradance of a light source

250 nm to 1020 nm

3.4% to 9.8%

Measurement by a working standard diode array spectroradiometer

On-site Calibration

123

13035-35090

Measurement of integrated absolute irradance of a light source

320 nm to 400 nm (UVA)

4.2%

Measurement by a working standard radiometer

On-site Calibration

124

13035-35091

Measurement of integrated absolute irradance of a light source

260 nm to 320 nm (UVB)

6.4%

Measurement by a working standard radiometer

On-site Calibration

125

13035-35098

Measurement of integrated absolute irradance of a light source

400 nm to 500 nm

3.6%

Measurement by a working standard radiometer

On-site Calibration

126

13034-33014

Solar simulator performance testing according to IEC 60904-9

400 nm to 1,100 nm for spectral match evaluation

0.006* to 0.02*

Measurements based on IEC 60904-9

On-site Calibration

127

13034-33014

Solar simulator performance testing according to IEC 60904-9

? 6,606 cm2 for designated test area of non-uniformity evaluation

0.10%

Measurements based on IEC 60904-9

On-site Calibration

128

13034-33014

Solar simulator performance testing according to IEC 60904-9

Data sampling time of DUT ? 2 ms for instability evaluation

0.05%

Measurements based on IEC 60904-9

On-site Calibration

129

13015-11011

Luminous intensity of an incandescent lamp

10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K

0.79%

Substitution method against two national reference standard lamps

Calibration Fee

130

13016-11011

Luminous intensity of an incandescent lamp

10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K

0.93%

Substitution method against two working standard lamps

Calibration Fee

131

13015-11021

Illuminance responsivity of a photo-detector head

12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K

0.70%

Source-based measurements using national reference standard lamps applying the inverse-square law.

Calibration Fee

132

13016-11021

Illuminance responsivity of a photo-detector head

12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K

0.80%

Source-based measurements using working standard lamps applying the inverse-square law.

Calibration Fee

133

13011-11021

Illuminance responsivity of an illuminance meter or a lux meter

2 lx ? ?? ? 10 000 lx 2 000 K ? CCT ? 3 400 K

Greater than or equal to 0.91% varied with illuminance level and UUC

Detector-based comparison against reference photometers

Calibration Fee

134

13010-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.66%

Detector-based method by using a gonio-photometer with a standard photometer

Calibration Fee

135

13015-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.80%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

136

13016-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.98%

Substitution method inside a 2-meter integrating sphere against two working standard lamps

Calibration Fee

137

13015-11032

Total luminous flux of a non-directional LED lamp

20 lm ? ?? ? 4 000 lm Full (non-directional), white

0.79%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

138

13016-11032

Total luminous flux of a non-directional LED lamp

20 lm ? ?? ? 4 000 lm Full (non-directional), white

1.0%

Substitution method inside a 2-meter integrating sphere against two working standard lamps

Calibration Fee

139

13015-11033

Total luminous flux of a linear fluorescent lamp

800 lm ? ?? ? 4 000 lm 2 600 K ? CCT ? 6 200 K

0.88%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

140

13011-11060

Luminance responsivity of a luminance meter

10 cd/m2 ? L? ? 10 000 cd/m2 2 000 K ? CCT ? 3 400 K

0.85%

Detector-based comparison against reference photometers

Calibration Fee

141

13011-11060

Luminance responsivity of a luminance meter

1 cd/m2 ? L? < 10 cd/m2 2 000 K ? CCT ? 3 400 K

1.2%

Detector-based comparison against reference photometers and a monitor detector

Calibration Fee

142

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

31.6% < %T ? 97.7% (0.01 ? OD < 0.5)

200 nm to 2000 nm: Greater than or equal to 0.48%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

143

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

10.0% < %T ? 31.6% (0.5 ? OD < 1.0)

200 nm to 2000: Greater than or equal to 0.57%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

144

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

1.00% < %T ? 10.00% (1.0 ? OD <2.0)

200 nm to 2000: Greater than or equal to 0.61%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

145

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

0.01% < %T ?1.00% (2.0 ? OD <4.0)

200 nm to 1800: Greater than or equal to 0.81%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

146

13025-14021

Diffuse spectral transmittance of a diffusely transmitting material

380 nm to 780 nm

For 20% haze plate, typical uncertainties in diffuse spectral transmittance are 0.13% to 0.18% throughout the spectrum

Direct measurement by single-beam primary reference spectrophtometer

Calibration Fee

147

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

148

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

149

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

150

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

151

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

152

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

153

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

154

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

155

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

156

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

157

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

158

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

159

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

380 nm to 400 nm: 2.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

160

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

405 nm to 460 nm: 1.6%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

161

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

465 nm to 780 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

162

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

163

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

164

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

165

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

166

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

167

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

168

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

169

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

170

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

171

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

172

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

173

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

174

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

380 nm to 400 nm: 2.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

175

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

405 nm to 460 nm: 1.6%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

176

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

465 nm to 780 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

177

13025-14061

Regular spectral reflectance

5 ? %R ? 100% 200 nm ? l ? 2 500 nm at 5 nm interval

0.32% to 0.42% for the wavelength range from 200 nm to 885 nm 1.0% to 1.4% for the wavelength range from 890 nm to 2 500 nm *Typical values for a high reflective aluminum mirror. Uncertainties for other materials are likely higher.

Absolute method by a double-subtractive double-beam spectrophotometer (Perkin-Elmer Lambda 1050) equipped with a variable-angle reflectance accessory (URA)

Calibration Fee

178

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

179

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

180

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

181

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

182

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

183

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

184

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

185

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

186

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

187

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

188

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

189

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

190

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

380 nm to 400 nm: 2.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

191

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

405 nm to 460 nm: 1.6%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

192

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

465 nm to 780 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

193

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

194

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

195

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

196

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

197

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

198

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

199

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

200

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

201

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

202

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

203

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

204

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

205

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

380 nm to 400 nm: 2.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

206

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

405 nm to 460 nm: 1.6%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

207

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

465 nm to 780 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

208

13025-14150

Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.)

200 nm to 2 000 nm

200 nm to 830 nm: 0.08 nm

Measurement by reference spectrometer

Calibration Fee

209

13025-14150

Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.)

200 nm to 2 000 nm

830 nm to 2 000 nm: 0.50 nm

Measurement by reference spectrometer

Calibration Fee

210

13025-14151

Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.)

200 nm to 2 000 nm

200 nm to 830 nm: 0.15 nm

Measurement by reference spectrometer with integrating sphere accessory

Calibration Fee

211

13025-14151

Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.)

200 nm to 2 000 nm

830 nm to 2 000 nm: 0.70 nm

Measurement by reference spectrometer with integrating sphere accessory

Calibration Fee

212

13025-16010

Surface color of general reflecting material, x, y, Y, u', v', L*, a*, b*

Varied be parameter of interest, please contact a laboratory staff for full details.

Varied by parameter of interest and color of UUC, please contact a laboratory staff for full details.

Spectral diffuse reflectance measurement by single-beam substitution method against a reference color standard

Calibration Fee

213

13025-16030

Transmitted color of a transmitting material

360 nm to 860 nm

Values and uncertainties vary depending on sample's color. Contact the Colorimetry Laboratory for details

Direct measurement by single-beam primary reference spectrophtometer

Calibration Fee

214

13025-16070

Haze and luminous transmittance of a diffusely transmitting material

380 nm to 780 nm

For 20% haze plate, typical uncertainty in transmission haze is 0.16%.

Based on ASTM D1003 standard method

Calibration Fee

215

13025-16120

Whiteness and Tint

380 nm to 780 nm W greater than 40 and less than 5Y-280

1.5

Spectral diffuse reflectance measurement by single-beam substitution method against a reference white diffuse reflectance standard

Calibration Fee

216

13038-12023

Absolute spectral irradiance responsivity of a scanning spectroradiometer with irradiance probe

250 nm to 2 500 nm

Greater than or equal to 2.5%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

217

13039-12023

Absolute spectral irradiance responsivity of a diode array spectroradiometer with irradiance probe

250 nm to 2 500 nm

Greater than or equal to 2.4%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

218

13038-12030

Relative spectral radiance responsivity of a scanning spectroradiometer with radiance probe

250 nm to 2 500 nm

Greater than or equal to 2.5%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

219

13039-12030

Relative spectral radiance responsivity of a diode array spectroradiometer with radiance probe

250 nm to 2 500 nm

Greater than or equal to 2.4%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

220

13034-12071

Solar irradiance responsivity of a solar power meter

400 nm to 1 100 nm depending on UUC

4.2%

Comparison against a working standard spectroradiometer applying 300 W Xe arc lamp with airmass 1.5 Global filter

Calibration Fee

221

13034-12090

Intregrated irradiance responsivity of broadband radiometer

315 nm to 400 nm (UVA )

4.4%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

222

13034-12091

Intregrated irradiance responsivity of broadband radiometer

280 nm to 315 nm (UVB)

6.3%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

223

13034-12092

Intregrated irradiance responsivity of broadband radiometer

280 nm to 400 nm (UVA+UVB)

6.0%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

224

13034-12099

Intregrated irradiance responsivity of broadband radiometer

Arbitary range within 280 nm to 1 020 nm

Varied by selected range

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

225

13034-12110

Intregrated irradiance responsivity of narrowband radiometer

365 nm ? 10 nm

4.4%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

226

13034-12113

Intregrated irradiance responsivity of narrowband radiometer

254 nm ? 10 nm

10%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

227

13034-12119

Intregrated irradiance responsivity of narrowband radiometer

Arbitary range within 280 nm to 1 020 nm

Varied by selected range

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

228

13038-12120

Responsivity of PAR radiometer

400 nm to 700 nm

2.0%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

229

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

250 nm ? l ? 320 nm, BW = 5 nm

4.3% to 7.6%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

230

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

320 nm < l ? 400 nm, BW = 5 nm

2.5% to 3.8%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

231

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

400 nm < l ? 1 100 nm, BW = 5 nm

2.2% to 2.3%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

232

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

1 100 nm < l ? 1 800 nm, BW = 5 nm

2.7% to 2.9%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

233

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

1 800 nm < l ? 2 300 nm, BW = 10 nm

3.6% to 4.5%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

234

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

2 300 nm < l ? 2 500 nm, BW = 10 nm

4.9% to 10%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

235

13037-13012

Spectral irradiance of a duterium lamp

240 nm to 400 nm

3.8% to 10%

By scanning measurements against a spectral irradiance reference national standard lamp using a double subtractive monochromator

Calibration Fee

236

13034-13013

Measurement of absolute / relative spectral irradance of a light source

250 nm to 1 020 nm

Greater than or equal to 2.4%

Measurement by a working standard diode array spectroradiometer

Calibration Fee

237

13038-15030

Correlated colour temperature response of a filter radiometer/Spectrometer

3 000 K to 3 200 K

20 K or 0.70%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

238

13035-15090

Measurement of integrated absolute irradance of a light source

320 nm to 400 nm (UVA)

4.2%

Measurement by a working standard radiometer

Calibration Fee

239

13035-15091

Measurement of integrated absolute irradance of a light source

260 nm to 320 nm (UVB)

6.4%

Measurement by a working standard radiometer

Calibration Fee

240

13035-15098

Measurement of integrated absolute irradance of a light source

400 nm to 500 nm

3.6%

Measurement by a working standard radiometer

Calibration Fee

241

13032-12900

Measurement of linearity response of a photo-detector

At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm

200 nm to 400 nm: ? 0.1% 400 nm to 2 500 nm: ? 0.05%

Flux superposition method

Calibration Fee

242

13032-12910

Measurement of uniformity of response of a photo-detector

At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm

201 nm to 400 nm: ? 0.2% 400 nm to 2 500 nm: ? 0.1%

x-y linear scan

Calibration Fee

243

13045-11012

Luminous intensity of a single packaged LED

0.1 cd to 100 cd

2.3% to 3.4% varied with LED color

Spectroradiometric method against two spectral irradiance standard lamps for CIE condition A and B

Calibration Fee

244

13045-11032

Luminous flux of a single-packaged LED

0.1 lm to 200 lm

1.8% to 2.3% varied with LED color

Absolute integrating sphere method against two spectral irradiance standard lamps

Calibration Fee

245

13045-13051

Total spectral radiant flux of a tungsten lamp

1.49 mW/nm ? ?e ? 35.8 mW/nm 360 nm ? l ? 830 nm BW = 1 nm

3.1% to 6.2%

Absolute integrating sphere method against spectral irradiance standard lamps

Calibration Fee

246

13041-15010

Distribution temperature of tungsten lamp

2 600 K to 3 200 K

20 K* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

247

13041-15021

Correlated colour temperature of general light source

2 600 K to 6 600 K

21 K* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

248

13041-15040

Chromaticity coordinate of general light source

0 to 0.9 at approximately 1 nm bandwidth

U(x)=0.001 7* U(y)=0.000 6* U(u)=0.001 0* U(v)=0.000 3* U(u')=0.001 0* U(v')=0.000 5* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

249

13045-15050

Chromaticity response of a colorimeter

x=0.426 9 y=0.400 7 u'=0.245 5 v'=0.518 5

0.004 0 0.002 0 0.002 3 0.001 2

Source-based calibration against a standard lamp

Calibration Fee

250

13034-33013

Measurement of absolute / relative spectral irradance of a light source

250 nm to 1020 nm

3.4% to 9.8%

Measurement by a working standard diode array spectroradiometer

On-site Calibration

251

13035-35090

Measurement of integrated absolute irradance of a light source

320 nm to 400 nm (UVA)

4.2%

Measurement by a working standard radiometer

On-site Calibration

252

13035-35091

Measurement of integrated absolute irradance of a light source

260 nm to 320 nm (UVB)

6.4%

Measurement by a working standard radiometer

On-site Calibration

253

13035-35098

Measurement of integrated absolute irradance of a light source

400 nm to 500 nm

3.6%

Measurement by a working standard radiometer

On-site Calibration

254

13034-33014

Solar simulator performance testing according to IEC 60904-9

400 nm to 1,100 nm for spectral match evaluation

0.006* to 0.02*

Measurements based on IEC 60904-9

On-site Calibration

255

13034-33014

Solar simulator performance testing according to IEC 60904-9

? 6,606 cm2 for designated test area of non-uniformity evaluation

0.10%

Measurements based on IEC 60904-9

On-site Calibration

256

13034-33014

Solar simulator performance testing according to IEC 60904-9

Data sampling time of DUT ? 2 ms for instability evaluation

0.05%

Measurements based on IEC 60904-9

On-site Calibration

257

13015-11011

Luminous intensity of an incandescent lamp

10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K

0.79%

Substitution method against two national reference standard lamps

Calibration Fee

258

13016-11011

Luminous intensity of an incandescent lamp

10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K

0.93%

Substitution method against two working standard lamps

Calibration Fee

259

13015-11021

Illuminance responsivity of a photo-detector head

12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K

0.70%

Source-based measurements using national reference standard lamps applying the inverse-square law.

Calibration Fee

260

13016-11021

Illuminance responsivity of a photo-detector head

12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K

0.80%

Source-based measurements using working standard lamps applying the inverse-square law.

Calibration Fee

261

13011-11021

Illuminance responsivity of an illuminance meter or a lux meter

2 lx ? ?? ? 10 000 lx 2 000 K ? CCT ? 3 400 K

Greater than or equal to 0.91% varied with illuminance level and UUC

Detector-based comparison against reference photometers

Calibration Fee

262

13010-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.66%

Detector-based method by using a gonio-photometer with a standard photometer

Calibration Fee

263

13015-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.80%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

264

13016-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.98%

Substitution method inside a 2-meter integrating sphere against two working standard lamps

Calibration Fee

265

13015-11032

Total luminous flux of a non-directional LED lamp

20 lm ? ?? ? 4 000 lm Full (non-directional), white

0.79%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

266

13016-11032

Total luminous flux of a non-directional LED lamp

20 lm ? ?? ? 4 000 lm Full (non-directional), white

1.0%

Substitution method inside a 2-meter integrating sphere against two working standard lamps

Calibration Fee

267

13015-11033

Total luminous flux of a linear fluorescent lamp

800 lm ? ?? ? 4 000 lm 2 600 K ? CCT ? 6 200 K

0.88%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

268

13011-11060

Luminance responsivity of a luminance meter

10 cd/m2 ? L? ? 10 000 cd/m2 2 000 K ? CCT ? 3 400 K

0.85%

Detector-based comparison against reference photometers

Calibration Fee

269

13011-11060

Luminance responsivity of a luminance meter

1 cd/m2 ? L? < 10 cd/m2 2 000 K ? CCT ? 3 400 K

1.2%

Detector-based comparison against reference photometers and a monitor detector

Calibration Fee

270

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

31.6% < %T ? 97.7% (0.01 ? OD < 0.5)

200 nm to 2000 nm: Greater than or equal to 0.48%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

271

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

10.0% < %T ? 31.6% (0.5 ? OD < 1.0)

200 nm to 2000: Greater than or equal to 0.57%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

272

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

1.00% < %T ? 10.00% (1.0 ? OD <2.0)

200 nm to 2000: Greater than or equal to 0.61%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

273

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

0.01% < %T ?1.00% (2.0 ? OD <4.0)

200 nm to 1800: Greater than or equal to 0.81%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

274

13025-14021

Diffuse spectral transmittance of a diffusely transmitting material

380 nm to 780 nm

For 20% haze plate, typical uncertainties in diffuse spectral transmittance are 0.13% to 0.18% throughout the spectrum

Direct measurement by single-beam primary reference spectrophtometer

Calibration Fee

275

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

276

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

277

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

278

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

279

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

280

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

281

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

282

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

283

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

284

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

285

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

286

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

287

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

380 nm to 400 nm: 2.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

288

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

405 nm to 460 nm: 1.6%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

289

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

465 nm to 780 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

290

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

291

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

292

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

293

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

294

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

295

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

296

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

297

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

298

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

299

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

300

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

301

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

302

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

380 nm to 400 nm: 2.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

303

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

405 nm to 460 nm: 1.6%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

304

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

465 nm to 780 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

305

13025-14061

Regular spectral reflectance

5 ? %R ? 100% 200 nm ? l ? 2 500 nm at 5 nm interval

0.32% to 0.42% for the wavelength range from 200 nm to 885 nm 1.0% to 1.4% for the wavelength range from 890 nm to 2 500 nm *Typical values for a high reflective aluminum mirror. Uncertainties for other materials are likely higher.

Absolute method by a double-subtractive double-beam spectrophotometer (Perkin-Elmer Lambda 1050) equipped with a variable-angle reflectance accessory (URA)

Calibration Fee

306

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

307

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

308

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

309

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

310

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

311

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

312

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

313

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

314

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

315

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

316

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

317

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

318

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

380 nm to 400 nm: 2.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

319

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

405 nm to 460 nm: 1.6%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

320

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

465 nm to 780 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

321

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

322

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

323

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

324

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

325

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

326

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

327

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

328

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

329

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

330

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

331

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

332

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

333

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

380 nm to 400 nm: 2.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

334

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

405 nm to 460 nm: 1.6%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

335

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

465 nm to 780 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

336

13025-14150

Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.)

200 nm to 2 000 nm

200 nm to 830 nm: 0.08 nm

Measurement by reference spectrometer

Calibration Fee

337

13025-14150

Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.)

200 nm to 2 000 nm

830 nm to 2 000 nm: 0.50 nm

Measurement by reference spectrometer

Calibration Fee

338

13025-14151

Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.)

200 nm to 2 000 nm

200 nm to 830 nm: 0.15 nm

Measurement by reference spectrometer with integrating sphere accessory

Calibration Fee

339

13025-14151

Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.)

200 nm to 2 000 nm

830 nm to 2 000 nm: 0.70 nm

Measurement by reference spectrometer with integrating sphere accessory

Calibration Fee

340

13025-16010

Surface color of general reflecting material, x, y, Y, u', v', L*, a*, b*

Varied be parameter of interest, please contact a laboratory staff for full details.

Varied by parameter of interest and color of UUC, please contact a laboratory staff for full details.

Spectral diffuse reflectance measurement by single-beam substitution method against a reference color standard

Calibration Fee

341

13025-16030

Transmitted color of a transmitting material

360 nm to 860 nm

Values and uncertainties vary depending on sample's color. Contact the Colorimetry Laboratory for details

Direct measurement by single-beam primary reference spectrophtometer

Calibration Fee

342

13025-16070

Haze and luminous transmittance of a diffusely transmitting material

380 nm to 780 nm

For 20% haze plate, typical uncertainty in transmission haze is 0.16%.

Based on ASTM D1003 standard method

Calibration Fee

343

13025-16120

Whiteness and Tint

380 nm to 780 nm W greater than 40 and less than 5Y-280

1.5

Spectral diffuse reflectance measurement by single-beam substitution method against a reference white diffuse reflectance standard

Calibration Fee

344

13038-12023

Absolute spectral irradiance responsivity of a scanning spectroradiometer with irradiance probe

250 nm to 2 500 nm

Greater than or equal to 2.5%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

345

13039-12023

Absolute spectral irradiance responsivity of a diode array spectroradiometer with irradiance probe

250 nm to 2 500 nm

Greater than or equal to 2.4%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

346

13038-12030

Relative spectral radiance responsivity of a scanning spectroradiometer with radiance probe

250 nm to 2 500 nm

Greater than or equal to 2.5%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

347

13039-12030

Relative spectral radiance responsivity of a diode array spectroradiometer with radiance probe

250 nm to 2 500 nm

Greater than or equal to 2.4%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

348

13034-12071

Solar irradiance responsivity of a solar power meter

400 nm to 1 100 nm depending on UUC

4.2%

Comparison against a working standard spectroradiometer applying 300 W Xe arc lamp with airmass 1.5 Global filter

Calibration Fee

349

13034-12090

Intregrated irradiance responsivity of broadband radiometer

315 nm to 400 nm (UVA )

4.4%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

350

13034-12091

Intregrated irradiance responsivity of broadband radiometer

280 nm to 315 nm (UVB)

6.3%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

351

13034-12092

Intregrated irradiance responsivity of broadband radiometer

280 nm to 400 nm (UVA+UVB)

6.0%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

352

13034-12099

Intregrated irradiance responsivity of broadband radiometer

Arbitary range within 280 nm to 1 020 nm

Varied by selected range

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

353

13034-12110

Intregrated irradiance responsivity of narrowband radiometer

365 nm ? 10 nm

4.4%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

354

13034-12113

Intregrated irradiance responsivity of narrowband radiometer

254 nm ? 10 nm

10%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

355

13034-12119

Intregrated irradiance responsivity of narrowband radiometer

Arbitary range within 280 nm to 1 020 nm

Varied by selected range

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

356

13038-12120

Responsivity of PAR radiometer

400 nm to 700 nm

2.0%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

357

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

250 nm ? l ? 320 nm, BW = 5 nm

4.3% to 7.6%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

358

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

320 nm < l ? 400 nm, BW = 5 nm

2.5% to 3.8%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

359

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

400 nm < l ? 1 100 nm, BW = 5 nm

2.2% to 2.3%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

360

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

1 100 nm < l ? 1 800 nm, BW = 5 nm

2.7% to 2.9%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

361

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

1 800 nm < l ? 2 300 nm, BW = 10 nm

3.6% to 4.5%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

362

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

2 300 nm < l ? 2 500 nm, BW = 10 nm

4.9% to 10%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

363

13037-13012

Spectral irradiance of a duterium lamp

240 nm to 400 nm

3.8% to 10%

By scanning measurements against a spectral irradiance reference national standard lamp using a double subtractive monochromator

Calibration Fee

364

13034-13013

Measurement of absolute / relative spectral irradance of a light source

250 nm to 1 020 nm

Greater than or equal to 2.4%

Measurement by a working standard diode array spectroradiometer

Calibration Fee

365

13038-15030

Correlated colour temperature response of a filter radiometer/Spectrometer

3 000 K to 3 200 K

20 K or 0.70%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

366

13035-15090

Measurement of integrated absolute irradance of a light source

320 nm to 400 nm (UVA)

4.2%

Measurement by a working standard radiometer

Calibration Fee

367

13035-15091

Measurement of integrated absolute irradance of a light source

260 nm to 320 nm (UVB)

6.4%

Measurement by a working standard radiometer

Calibration Fee

368

13035-15098

Measurement of integrated absolute irradance of a light source

400 nm to 500 nm

3.6%

Measurement by a working standard radiometer

Calibration Fee

369

13032-12900

Measurement of linearity response of a photo-detector

At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm

200 nm to 400 nm: ? 0.1% 400 nm to 2 500 nm: ? 0.05%

Flux superposition method

Calibration Fee

370

13032-12910

Measurement of uniformity of response of a photo-detector

At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm

201 nm to 400 nm: ? 0.2% 400 nm to 2 500 nm: ? 0.1%

x-y linear scan

Calibration Fee

371

13045-11012

Luminous intensity of a single packaged LED

0.1 cd to 100 cd

2.3% to 3.4% varied with LED color

Spectroradiometric method against two spectral irradiance standard lamps for CIE condition A and B

Calibration Fee

372

13045-11032

Luminous flux of a single-packaged LED

0.1 lm to 200 lm

1.8% to 2.3% varied with LED color

Absolute integrating sphere method against two spectral irradiance standard lamps

Calibration Fee

373

13045-13051

Total spectral radiant flux of a tungsten lamp

1.49 mW/nm ? ?e ? 35.8 mW/nm 360 nm ? l ? 830 nm BW = 1 nm

3.1% to 6.2%

Absolute integrating sphere method against spectral irradiance standard lamps

Calibration Fee

374

13041-15010

Distribution temperature of tungsten lamp

2 600 K to 3 200 K

20 K* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

375

13041-15021

Correlated colour temperature of general light source

2 600 K to 6 600 K

21 K* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

376

13041-15040

Chromaticity coordinate of general light source

0 to 0.9 at approximately 1 nm bandwidth

U(x)=0.001 7* U(y)=0.000 6* U(u)=0.001 0* U(v)=0.000 3* U(u')=0.001 0* U(v')=0.000 5* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

377

13045-15050

Chromaticity response of a colorimeter

x=0.426 9 y=0.400 7 u'=0.245 5 v'=0.518 5

0.004 0 0.002 0 0.002 3 0.001 2

Source-based calibration against a standard lamp

Calibration Fee

378

13034-33013

Measurement of absolute / relative spectral irradance of a light source

250 nm to 1020 nm

3.4% to 9.8%

Measurement by a working standard diode array spectroradiometer

On-site Calibration

379

13035-35090

Measurement of integrated absolute irradance of a light source

320 nm to 400 nm (UVA)

4.2%

Measurement by a working standard radiometer

On-site Calibration

380

13035-35091

Measurement of integrated absolute irradance of a light source

260 nm to 320 nm (UVB)

6.4%

Measurement by a working standard radiometer

On-site Calibration

381

13035-35098

Measurement of integrated absolute irradance of a light source

400 nm to 500 nm

3.6%

Measurement by a working standard radiometer

On-site Calibration

382

13034-33014

Solar simulator performance testing according to IEC 60904-9

400 nm to 1,100 nm for spectral match evaluation

0.006* to 0.02*

Measurements based on IEC 60904-9

On-site Calibration

383

13034-33014

Solar simulator performance testing according to IEC 60904-9

? 6,606 cm2 for designated test area of non-uniformity evaluation

0.10%

Measurements based on IEC 60904-9

On-site Calibration

384

13034-33014

Solar simulator performance testing according to IEC 60904-9

Data sampling time of DUT ? 2 ms for instability evaluation

0.05%

Measurements based on IEC 60904-9

On-site Calibration

385

13015-11011

Luminous intensity of an incandescent lamp

10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K

0.79%

Substitution method against two national reference standard lamps

Calibration Fee

386

13016-11011

Luminous intensity of an incandescent lamp

10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K

0.93%

Substitution method against two working standard lamps

Calibration Fee

387

13015-11021

Illuminance responsivity of a photo-detector head

12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K

0.70%

Source-based measurements using national reference standard lamps applying the inverse-square law.

Calibration Fee

388

13016-11021

Illuminance responsivity of a photo-detector head

12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K

0.80%

Source-based measurements using working standard lamps applying the inverse-square law.

Calibration Fee

389

13011-11021

Illuminance responsivity of an illuminance meter or a lux meter

2 lx ? ?? ? 10 000 lx 2 000 K ? CCT ? 3 400 K

Greater than or equal to 0.91% varied with illuminance level and UUC

Detector-based comparison against reference photometers

Calibration Fee

390

13010-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.66%

Detector-based method by using a gonio-photometer with a standard photometer

Calibration Fee

391

13015-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.80%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

392

13016-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.98%

Substitution method inside a 2-meter integrating sphere against two working standard lamps

Calibration Fee

393

13015-11032

Total luminous flux of a non-directional LED lamp

20 lm ? ?? ? 4 000 lm Full (non-directional), white

0.79%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

394

13016-11032

Total luminous flux of a non-directional LED lamp

20 lm ? ?? ? 4 000 lm Full (non-directional), white

1.0%

Substitution method inside a 2-meter integrating sphere against two working standard lamps

Calibration Fee

395

13015-11033

Total luminous flux of a linear fluorescent lamp

800 lm ? ?? ? 4 000 lm 2 600 K ? CCT ? 6 200 K

0.88%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

396

13011-11060

Luminance responsivity of a luminance meter

10 cd/m2 ? L? ? 10 000 cd/m2 2 000 K ? CCT ? 3 400 K

0.85%

Detector-based comparison against reference photometers

Calibration Fee

397

13011-11060

Luminance responsivity of a luminance meter

1 cd/m2 ? L? < 10 cd/m2 2 000 K ? CCT ? 3 400 K

1.2%

Detector-based comparison against reference photometers and a monitor detector

Calibration Fee

398

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

31.6% < %T ? 97.7% (0.01 ? OD < 0.5)

200 nm to 2000 nm: Greater than or equal to 0.48%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

399

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

10.0% < %T ? 31.6% (0.5 ? OD < 1.0)

200 nm to 2000: Greater than or equal to 0.57%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

400

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

1.00% < %T ? 10.00% (1.0 ? OD <2.0)

200 nm to 2000: Greater than or equal to 0.61%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

401

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

0.01% < %T ?1.00% (2.0 ? OD <4.0)

200 nm to 1800: Greater than or equal to 0.81%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

402

13025-14021

Diffuse spectral transmittance of a diffusely transmitting material

380 nm to 780 nm

For 20% haze plate, typical uncertainties in diffuse spectral transmittance are 0.13% to 0.18% throughout the spectrum

Direct measurement by single-beam primary reference spectrophtometer

Calibration Fee

403

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

404

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

405

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

406

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

407

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

408

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

409

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

410

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

411

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

412

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

413

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

414

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

415

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

380 nm to 400 nm: 2.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

416

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

405 nm to 460 nm: 1.6%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

417

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

465 nm to 780 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

418

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

419

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

420

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

421

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

422

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

423

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

424

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

425

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

426

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

427

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

428

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

429

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

430

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

380 nm to 400 nm: 2.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

431

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

405 nm to 460 nm: 1.6%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

432

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

465 nm to 780 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

433

13025-14061

Regular spectral reflectance

5 ? %R ? 100% 200 nm ? l ? 2 500 nm at 5 nm interval

0.32% to 0.42% for the wavelength range from 200 nm to 885 nm 1.0% to 1.4% for the wavelength range from 890 nm to 2 500 nm *Typical values for a high reflective aluminum mirror. Uncertainties for other materials are likely higher.

Absolute method by a double-subtractive double-beam spectrophotometer (Perkin-Elmer Lambda 1050) equipped with a variable-angle reflectance accessory (URA)

Calibration Fee

434

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

435

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

436

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

437

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

438

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

439

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

440

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

441

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

442

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

443

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

444

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

445

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

446

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

380 nm to 400 nm: 2.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

447

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

405 nm to 460 nm: 1.6%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

448

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

465 nm to 780 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

449

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

450

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

451

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

452

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

453

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

454

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

455

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

456

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

457

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

458

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

459

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

460

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

461

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

380 nm to 400 nm: 2.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

462

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

405 nm to 460 nm: 1.6%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

463

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

465 nm to 780 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

464

13025-14150

Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.)

200 nm to 2 000 nm

200 nm to 830 nm: 0.08 nm

Measurement by reference spectrometer

Calibration Fee

465

13025-14150

Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.)

200 nm to 2 000 nm

830 nm to 2 000 nm: 0.50 nm

Measurement by reference spectrometer

Calibration Fee

466

13025-14151

Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.)

200 nm to 2 000 nm

200 nm to 830 nm: 0.15 nm

Measurement by reference spectrometer with integrating sphere accessory

Calibration Fee

467

13025-14151

Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.)

200 nm to 2 000 nm

830 nm to 2 000 nm: 0.70 nm

Measurement by reference spectrometer with integrating sphere accessory

Calibration Fee

468

13025-16010

Surface color of general reflecting material, x, y, Y, u', v', L*, a*, b*

Varied be parameter of interest, please contact a laboratory staff for full details.

Varied by parameter of interest and color of UUC, please contact a laboratory staff for full details.

Spectral diffuse reflectance measurement by single-beam substitution method against a reference color standard

Calibration Fee

469

13025-16030

Transmitted color of a transmitting material

360 nm to 860 nm

Values and uncertainties vary depending on sample's color. Contact the Colorimetry Laboratory for details

Direct measurement by single-beam primary reference spectrophtometer

Calibration Fee

470

13025-16070

Haze and luminous transmittance of a diffusely transmitting material

380 nm to 780 nm

For 20% haze plate, typical uncertainty in transmission haze is 0.16%.

Based on ASTM D1003 standard method

Calibration Fee

471

13025-16120

Whiteness and Tint

380 nm to 780 nm W greater than 40 and less than 5Y-280

1.5

Spectral diffuse reflectance measurement by single-beam substitution method against a reference white diffuse reflectance standard

Calibration Fee

472

13038-12023

Absolute spectral irradiance responsivity of a scanning spectroradiometer with irradiance probe

250 nm to 2 500 nm

Greater than or equal to 2.5%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

473

13039-12023

Absolute spectral irradiance responsivity of a diode array spectroradiometer with irradiance probe

250 nm to 2 500 nm

Greater than or equal to 2.4%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

474

13038-12030

Relative spectral radiance responsivity of a scanning spectroradiometer with radiance probe

250 nm to 2 500 nm

Greater than or equal to 2.5%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

475

13039-12030

Relative spectral radiance responsivity of a diode array spectroradiometer with radiance probe

250 nm to 2 500 nm

Greater than or equal to 2.4%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

476

13034-12071

Solar irradiance responsivity of a solar power meter

400 nm to 1 100 nm depending on UUC

4.2%

Comparison against a working standard spectroradiometer applying 300 W Xe arc lamp with airmass 1.5 Global filter

Calibration Fee

477

13034-12090

Intregrated irradiance responsivity of broadband radiometer

315 nm to 400 nm (UVA )

4.4%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

478

13034-12091

Intregrated irradiance responsivity of broadband radiometer

280 nm to 315 nm (UVB)

6.3%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

479

13034-12092

Intregrated irradiance responsivity of broadband radiometer

280 nm to 400 nm (UVA+UVB)

6.0%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

480

13034-12099

Intregrated irradiance responsivity of broadband radiometer

Arbitary range within 280 nm to 1 020 nm

Varied by selected range

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

481

13034-12110

Intregrated irradiance responsivity of narrowband radiometer

365 nm ? 10 nm

4.4%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

482

13034-12113

Intregrated irradiance responsivity of narrowband radiometer

254 nm ? 10 nm

10%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

483

13034-12119

Intregrated irradiance responsivity of narrowband radiometer

Arbitary range within 280 nm to 1 020 nm

Varied by selected range

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

484

13038-12120

Responsivity of PAR radiometer

400 nm to 700 nm

2.0%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

485

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

250 nm ? l ? 320 nm, BW = 5 nm

4.3% to 7.6%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

486

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

320 nm < l ? 400 nm, BW = 5 nm

2.5% to 3.8%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

487

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

400 nm < l ? 1 100 nm, BW = 5 nm

2.2% to 2.3%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

488

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

1 100 nm < l ? 1 800 nm, BW = 5 nm

2.7% to 2.9%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

489

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

1 800 nm < l ? 2 300 nm, BW = 10 nm

3.6% to 4.5%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

490

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

2 300 nm < l ? 2 500 nm, BW = 10 nm

4.9% to 10%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

491

13037-13012

Spectral irradiance of a duterium lamp

240 nm to 400 nm

3.8% to 10%

By scanning measurements against a spectral irradiance reference national standard lamp using a double subtractive monochromator

Calibration Fee

492

13034-13013

Measurement of absolute / relative spectral irradance of a light source

250 nm to 1 020 nm

Greater than or equal to 2.4%

Measurement by a working standard diode array spectroradiometer

Calibration Fee

493

13038-15030

Correlated colour temperature response of a filter radiometer/Spectrometer

3 000 K to 3 200 K

20 K or 0.70%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

494

13035-15090

Measurement of integrated absolute irradance of a light source

320 nm to 400 nm (UVA)

4.2%

Measurement by a working standard radiometer

Calibration Fee

495

13035-15091

Measurement of integrated absolute irradance of a light source

260 nm to 320 nm (UVB)

6.4%

Measurement by a working standard radiometer

Calibration Fee

496

13035-15098

Measurement of integrated absolute irradance of a light source

400 nm to 500 nm

3.6%

Measurement by a working standard radiometer

Calibration Fee

497

13032-12900

Measurement of linearity response of a photo-detector

At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm

200 nm to 400 nm: ? 0.1% 400 nm to 2 500 nm: ? 0.05%

Flux superposition method

Calibration Fee

498

13032-12910

Measurement of uniformity of response of a photo-detector

At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm

201 nm to 400 nm: ? 0.2% 400 nm to 2 500 nm: ? 0.1%

x-y linear scan

Calibration Fee

499

13045-11012

Luminous intensity of a single packaged LED

0.1 cd to 100 cd

2.3% to 3.4% varied with LED color

Spectroradiometric method against two spectral irradiance standard lamps for CIE condition A and B

Calibration Fee

500

13045-11032

Luminous flux of a single-packaged LED

0.1 lm to 200 lm

1.8% to 2.3% varied with LED color

Absolute integrating sphere method against two spectral irradiance standard lamps

Calibration Fee

501

13045-13051

Total spectral radiant flux of a tungsten lamp

1.49 mW/nm ? ?e ? 35.8 mW/nm 360 nm ? l ? 830 nm BW = 1 nm

3.1% to 6.2%

Absolute integrating sphere method against spectral irradiance standard lamps

Calibration Fee

502

13041-15010

Distribution temperature of tungsten lamp

2 600 K to 3 200 K

20 K* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

503

13041-15021

Correlated colour temperature of general light source

2 600 K to 6 600 K

21 K* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

504

13041-15040

Chromaticity coordinate of general light source

0 to 0.9 at approximately 1 nm bandwidth

U(x)=0.001 7* U(y)=0.000 6* U(u)=0.001 0* U(v)=0.000 3* U(u')=0.001 0* U(v')=0.000 5* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

505

13045-15050

Chromaticity response of a colorimeter

x=0.426 9 y=0.400 7 u'=0.245 5 v'=0.518 5

0.004 0 0.002 0 0.002 3 0.001 2

Source-based calibration against a standard lamp

Calibration Fee

506

13034-33013

Measurement of absolute / relative spectral irradance of a light source

250 nm to 1020 nm

3.4% to 9.8%

Measurement by a working standard diode array spectroradiometer

On-site Calibration

507

13035-35090

Measurement of integrated absolute irradance of a light source

320 nm to 400 nm (UVA)

4.2%

Measurement by a working standard radiometer

On-site Calibration

508

13035-35091

Measurement of integrated absolute irradance of a light source

260 nm to 320 nm (UVB)

6.4%

Measurement by a working standard radiometer

On-site Calibration

509

13035-35098

Measurement of integrated absolute irradance of a light source

400 nm to 500 nm

3.6%

Measurement by a working standard radiometer

On-site Calibration

510

13034-33014

Solar simulator performance testing according to IEC 60904-9

400 nm to 1,100 nm for spectral match evaluation

0.006* to 0.02*

Measurements based on IEC 60904-9

On-site Calibration

511

13034-33014

Solar simulator performance testing according to IEC 60904-9

? 6,606 cm2 for designated test area of non-uniformity evaluation

0.10%

Measurements based on IEC 60904-9

On-site Calibration

512

13034-33014

Solar simulator performance testing according to IEC 60904-9

Data sampling time of DUT ? 2 ms for instability evaluation

0.05%

Measurements based on IEC 60904-9

On-site Calibration

513

13015-11011

Luminous intensity of an incandescent lamp

10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K

0.79%

Substitution method against two national reference standard lamps

Calibration Fee

514

13016-11011

Luminous intensity of an incandescent lamp

10 cd ? ?? ? 3 000 cd 2 000 K ? CCT ? 3 400 K

0.93%

Substitution method against two working standard lamps

Calibration Fee

515

13015-11021

Illuminance responsivity of a photo-detector head

12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K

0.70%

Source-based measurements using national reference standard lamps applying the inverse-square law.

Calibration Fee

516

13016-11021

Illuminance responsivity of a photo-detector head

12 lx ? ?? ? 300 lx 2 000 K ? CCT ? 3 400 K

0.80%

Source-based measurements using working standard lamps applying the inverse-square law.

Calibration Fee

517

13011-11021

Illuminance responsivity of an illuminance meter or a lux meter

2 lx ? ?? ? 10 000 lx 2 000 K ? CCT ? 3 400 K

Greater than or equal to 0.91% varied with illuminance level and UUC

Detector-based comparison against reference photometers

Calibration Fee

518

13010-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.66%

Detector-based method by using a gonio-photometer with a standard photometer

Calibration Fee

519

13015-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.80%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

520

13016-11031

Total luminous flux of an incandescent lamp

5 lm ? ?? ? 9 000 lm 2 000 K ? CCT ? 3 400 K

0.98%

Substitution method inside a 2-meter integrating sphere against two working standard lamps

Calibration Fee

521

13015-11032

Total luminous flux of a non-directional LED lamp

20 lm ? ?? ? 4 000 lm Full (non-directional), white

0.79%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

522

13016-11032

Total luminous flux of a non-directional LED lamp

20 lm ? ?? ? 4 000 lm Full (non-directional), white

1.0%

Substitution method inside a 2-meter integrating sphere against two working standard lamps

Calibration Fee

523

13015-11033

Total luminous flux of a linear fluorescent lamp

800 lm ? ?? ? 4 000 lm 2 600 K ? CCT ? 6 200 K

0.88%

Substitution method inside a 2-meter integrating sphere against two national reference standard lamps

Calibration Fee

524

13011-11060

Luminance responsivity of a luminance meter

10 cd/m2 ? L? ? 10 000 cd/m2 2 000 K ? CCT ? 3 400 K

0.85%

Detector-based comparison against reference photometers

Calibration Fee

525

13011-11060

Luminance responsivity of a luminance meter

1 cd/m2 ? L? < 10 cd/m2 2 000 K ? CCT ? 3 400 K

1.2%

Detector-based comparison against reference photometers and a monitor detector

Calibration Fee

526

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

31.6% < %T ? 97.7% (0.01 ? OD < 0.5)

200 nm to 2000 nm: Greater than or equal to 0.48%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

527

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

10.0% < %T ? 31.6% (0.5 ? OD < 1.0)

200 nm to 2000: Greater than or equal to 0.57%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

528

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

1.00% < %T ? 10.00% (1.0 ? OD <2.0)

200 nm to 2000: Greater than or equal to 0.61%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

529

13025-14011

Regular spectral transmittance of a transmitting filter or film. Measured transmittance values will be reported at 5 nm interval from 200 nm to 900 nm and 50 nm interval from 900 nm to 2 500 nm.

0.01% < %T ?1.00% (2.0 ? OD <4.0)

200 nm to 1800: Greater than or equal to 0.81%

By comparison against a reference ND filters using the reference spectrophotometer. Measurement uncertainties shown are percent relative of transmittance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

530

13025-14021

Diffuse spectral transmittance of a diffusely transmitting material

380 nm to 780 nm

For 20% haze plate, typical uncertainties in diffuse spectral transmittance are 0.13% to 0.18% throughout the spectrum

Direct measurement by single-beam primary reference spectrophtometer

Calibration Fee

531

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

532

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

533

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

534

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

535

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

536

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

537

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

538

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

539

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

540

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

541

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

542

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

543

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

380 nm to 400 nm: 2.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

544

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

405 nm to 460 nm: 1.6%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

545

13025-14051

Specular excluded spectral diffuse reflectance of a glossy color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

465 nm to 780 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

546

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

547

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

548

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

549

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

550

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

551

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

552

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

553

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

554

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

555

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

556

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

557

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

10% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

558

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

380 nm to 400 nm: 2.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

559

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

405 nm to 460 nm: 1.6%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

560

13026-14051

Specular excluded spectral diffuse reflectance of a matte color sample (8o:de) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

5% ? %R ? 10%

465 nm to 780 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

561

13025-14061

Regular spectral reflectance

5 ? %R ? 100% 200 nm ? l ? 2 500 nm at 5 nm interval

0.32% to 0.42% for the wavelength range from 200 nm to 885 nm 1.0% to 1.4% for the wavelength range from 890 nm to 2 500 nm *Typical values for a high reflective aluminum mirror. Uncertainties for other materials are likely higher.

Absolute method by a double-subtractive double-beam spectrophotometer (Perkin-Elmer Lambda 1050) equipped with a variable-angle reflectance accessory (URA)

Calibration Fee

562

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

563

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

564

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

565

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

566

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

567

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

568

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

569

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

570

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

571

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

572

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

573

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

574

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

380 nm to 400 nm: 2.1%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

575

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

405 nm to 460 nm: 1.6%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

576

13025-14071

Specular included spectral diffuse reflectance of a glossy color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

465 nm to 780 nm: 1.5%

By comparison against a reference glossy ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

577

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

578

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

579

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

75% < %R ? 100%

465 nm to 780 nm: 0.86%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

580

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

380 nm to 400 nm: 1.7%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

581

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

405 nm to 460 nm: 1.2%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

582

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

40% < %R ? 75%

465 nm to 780 nm:0.84%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

583

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

380 nm to 400 nm: 1.8%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

584

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

405 nm to 460 nm: 1.3%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

585

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

20% < %R ? 40%

465 nm to 780 nm: 0.90%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

586

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

380 nm to 400 nm: 1.9%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

587

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

405 nm to 460 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

588

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

7.5% < %R ? 20%

465 nm to 780 nm: 1.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

589

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

380 nm to 400 nm: 2.1%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

590

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

405 nm to 460 nm: 1.6%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

591

13026-14071

Specular included spectral diffuse reflectance of a matte color sample (8o:di) Measured diffused reflectance values will be reported at 5 nm interval from 380 nm to 780 nm.

2.5% ? %R ? 7.5%

465 nm to 780 nm: 1.5%

By comparison against a reference matte ceramic color standard using the reference spectrophotometer with integrating sphere accessory. Measurement uncertainties shown are percent relative of reflectance values and they are varied by wavelengths and dependent of UCC.

Calibration Fee

592

13025-14150

Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.)

200 nm to 2 000 nm

200 nm to 830 nm: 0.08 nm

Measurement by reference spectrometer

Calibration Fee

593

13025-14150

Wavelength calibration of a transmitted wavelength standard (holmium oxide solution, holmium oxide filter, dydimium oxide filter etc.)

200 nm to 2 000 nm

830 nm to 2 000 nm: 0.50 nm

Measurement by reference spectrometer

Calibration Fee

594

13025-14151

Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.)

200 nm to 2 000 nm

200 nm to 830 nm: 0.15 nm

Measurement by reference spectrometer with integrating sphere accessory

Calibration Fee

595

13025-14151

Wavelength calibration of a reflectance wavelength standard (holmium oxide doped ceramic tile etc.)

200 nm to 2 000 nm

830 nm to 2 000 nm: 0.70 nm

Measurement by reference spectrometer with integrating sphere accessory

Calibration Fee

596

13025-16010

Surface color of general reflecting material, x, y, Y, u', v', L*, a*, b*

Varied be parameter of interest, please contact a laboratory staff for full details.

Varied by parameter of interest and color of UUC, please contact a laboratory staff for full details.

Spectral diffuse reflectance measurement by single-beam substitution method against a reference color standard

Calibration Fee

597

13025-16030

Transmitted color of a transmitting material

360 nm to 860 nm

Values and uncertainties vary depending on sample's color. Contact the Colorimetry Laboratory for details

Direct measurement by single-beam primary reference spectrophtometer

Calibration Fee

598

13025-16070

Haze and luminous transmittance of a diffusely transmitting material

380 nm to 780 nm

For 20% haze plate, typical uncertainty in transmission haze is 0.16%.

Based on ASTM D1003 standard method

Calibration Fee

599

13025-16120

Whiteness and Tint

380 nm to 780 nm W greater than 40 and less than 5Y-280

1.5

Spectral diffuse reflectance measurement by single-beam substitution method against a reference white diffuse reflectance standard

Calibration Fee

600

13038-12023

Absolute spectral irradiance responsivity of a scanning spectroradiometer with irradiance probe

250 nm to 2 500 nm

Greater than or equal to 2.5%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

601

13039-12023

Absolute spectral irradiance responsivity of a diode array spectroradiometer with irradiance probe

250 nm to 2 500 nm

Greater than or equal to 2.4%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

602

13038-12030

Relative spectral radiance responsivity of a scanning spectroradiometer with radiance probe

250 nm to 2 500 nm

Greater than or equal to 2.5%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

603

13039-12030

Relative spectral radiance responsivity of a diode array spectroradiometer with radiance probe

250 nm to 2 500 nm

Greater than or equal to 2.4%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

604

13034-12071

Solar irradiance responsivity of a solar power meter

400 nm to 1 100 nm depending on UUC

4.2%

Comparison against a working standard spectroradiometer applying 300 W Xe arc lamp with airmass 1.5 Global filter

Calibration Fee

605

13034-12090

Intregrated irradiance responsivity of broadband radiometer

315 nm to 400 nm (UVA )

4.4%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

606

13034-12091

Intregrated irradiance responsivity of broadband radiometer

280 nm to 315 nm (UVB)

6.3%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

607

13034-12092

Intregrated irradiance responsivity of broadband radiometer

280 nm to 400 nm (UVA+UVB)

6.0%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

608

13034-12099

Intregrated irradiance responsivity of broadband radiometer

Arbitary range within 280 nm to 1 020 nm

Varied by selected range

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

609

13034-12110

Intregrated irradiance responsivity of narrowband radiometer

365 nm ? 10 nm

4.4%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

610

13034-12113

Intregrated irradiance responsivity of narrowband radiometer

254 nm ? 10 nm

10%

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

611

13034-12119

Intregrated irradiance responsivity of narrowband radiometer

Arbitary range within 280 nm to 1 020 nm

Varied by selected range

Source based calibration against a reference spectroradiometer using filtered 500 W Hg(Xe) as light source

Calibration Fee

612

13038-12120

Responsivity of PAR radiometer

400 nm to 700 nm

2.0%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

613

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

250 nm ? l ? 320 nm, BW = 5 nm

4.3% to 7.6%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

614

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

320 nm < l ? 400 nm, BW = 5 nm

2.5% to 3.8%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

615

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

400 nm < l ? 1 100 nm, BW = 5 nm

2.2% to 2.3%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

616

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

1 100 nm < l ? 1 800 nm, BW = 5 nm

2.7% to 2.9%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

617

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

1 800 nm < l ? 2 300 nm, BW = 10 nm

3.6% to 4.5%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

618

13037-13011

Spectral irradiance of a tungsten-halogen lamp (FEL and Ushio type)

2 300 nm < l ? 2 500 nm, BW = 10 nm

4.9% to 10%

By scanning measurements against a set of two spectral irradiance standard lamps using a double subtractive monochromator

Calibration Fee

619

13037-13012

Spectral irradiance of a duterium lamp

240 nm to 400 nm

3.8% to 10%

By scanning measurements against a spectral irradiance reference national standard lamp using a double subtractive monochromator

Calibration Fee

620

13034-13013

Measurement of absolute / relative spectral irradance of a light source

250 nm to 1 020 nm

Greater than or equal to 2.4%

Measurement by a working standard diode array spectroradiometer

Calibration Fee

621

13038-15030

Correlated colour temperature response of a filter radiometer/Spectrometer

3 000 K to 3 200 K

20 K or 0.70%

Direct measurement against a set of two FEL working standard lamps

Calibration Fee

622

13035-15090

Measurement of integrated absolute irradance of a light source

320 nm to 400 nm (UVA)

4.2%

Measurement by a working standard radiometer

Calibration Fee

623

13035-15091

Measurement of integrated absolute irradance of a light source

260 nm to 320 nm (UVB)

6.4%

Measurement by a working standard radiometer

Calibration Fee

624

13035-15098

Measurement of integrated absolute irradance of a light source

400 nm to 500 nm

3.6%

Measurement by a working standard radiometer

Calibration Fee

625

13032-12900

Measurement of linearity response of a photo-detector

At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm

200 nm to 400 nm: ? 0.1% 400 nm to 2 500 nm: ? 0.05%

Flux superposition method

Calibration Fee

626

13032-12910

Measurement of uniformity of response of a photo-detector

At discrete wavelengths from 200 nm to 2 500 nm Beam size ranges from 25 micron to 5 mm

201 nm to 400 nm: ? 0.2% 400 nm to 2 500 nm: ? 0.1%

x-y linear scan

Calibration Fee

627

13045-11012

Luminous intensity of a single packaged LED

0.1 cd to 100 cd

2.3% to 3.4% varied with LED color

Spectroradiometric method against two spectral irradiance standard lamps for CIE condition A and B

Calibration Fee

628

13045-11032

Luminous flux of a single-packaged LED

0.1 lm to 200 lm

1.8% to 2.3% varied with LED color

Absolute integrating sphere method against two spectral irradiance standard lamps

Calibration Fee

629

13045-13051

Total spectral radiant flux of a tungsten lamp

1.49 mW/nm ? ?e ? 35.8 mW/nm 360 nm ? l ? 830 nm BW = 1 nm

3.1% to 6.2%

Absolute integrating sphere method against spectral irradiance standard lamps

Calibration Fee

630

13041-15010

Distribution temperature of tungsten lamp

2 600 K to 3 200 K

20 K* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

631

13041-15021

Correlated colour temperature of general light source

2 600 K to 6 600 K

21 K* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

632

13041-15040

Chromaticity coordinate of general light source

0 to 0.9 at approximately 1 nm bandwidth

U(x)=0.001 7* U(y)=0.000 6* U(u)=0.001 0* U(v)=0.000 3* U(u')=0.001 0* U(v')=0.000 5* varied by UUC

Measurement by a working standard spectroradiometer

Calibration Fee

633

13045-15050

Chromaticity response of a colorimeter

x=0.426 9 y=0.400 7 u'=0.245 5 v'=0.518 5

0.004 0 0.002 0 0.002 3 0.001 2

Source-based calibration against a standard lamp

Calibration Fee

634

13034-33013

Measurement of absolute / relative spectral irradance of a light source

250 nm to 1020 nm

3.4% to 9.8%

Measurement by a working standard diode array spectroradiometer

On-site Calibration

635

13035-35090

Measurement of integrated absolute irradance of a light source

320 nm to 400 nm (UVA)

4.2%

Measurement by a working standard radiometer

On-site Calibration

636

13035-35091

Measurement of integrated absolute irradance of a light source

260 nm to 320 nm (UVB)

6.4%

Measurement by a working standard radiometer

On-site Calibration

637

13035-35098

Measurement of integrated absolute irradance of a light source

400 nm to 500 nm

3.6%

Measurement by a working standard radiometer

On-site Calibration

638

13034-33014

Solar simulator performance testing according to IEC 60904-9

400 nm to 1,100 nm for spectral match evaluation

0.006* to 0.02*

Measurements based on IEC 60904-9

On-site Calibration

639

13034-33014

Solar simulator performance testing according to IEC 60904-9

? 6,606 cm2 for designated test area of non-uniformity evaluation

0.10%

Measurements based on IEC 60904-9

On-site Calibration

640

13034-33014

Solar simulator performance testing according to IEC 60904-9

Data sampling time of DUT ? 2 ms for instability evaluation

0.05%

Measurements based on IEC 60904-9

On-site Calibration